Page 6
(5) Resistance (for 61-340 / 61-342)
Range Resolution
Accuracy
Over voltage
protection
400.0
Ω
*
1
0.1
Ω
±(1.2% r 5 digits)
4.000K
Ω
1
Ω
40.00K
Ω
10
Ω
400.0K
Ω
100
Ω
±(1.0% r 2 digits)
4.000M
Ω
1K
Ω
±(1.2% r 2 digits)
40.00M
Ω
*
2
10K
Ω
±(2.0% r 5 digits)
600Vrms
Open circuit Voltage:
0.44V approx.
*
1
< 5 digit of reading rolling.
*
2
< 2% of reading rolling.
(6) Diode Check and Continuity (for 61-340 / 61-342)
Range Resolution
Accuracy
Max. Test
Current
Max. Open
Circuit Voltage
1mV
Not
specified
*
<1mA, approx.
2.5V, approx.
Overload Protection:
Not specified
Continuity:
Built-in buzzer sounds when resistance is less than approximately 120
Ω
with a response
time of approximately 100 msec.
(7) Capacitance (for 61-340 / 61-342)
Range Resolution
Accuracy
Over voltage
Protection
40.00nF 10pF
REL
:±
(3% r 10 digits)
400.0nF 100pF
4.000
μ
F 1nF
40.00
μ
F 10nF
±(3% r 5 digits)
400.0
μ
F 100nF
4000
μ
F
*
1
10
μ
F
±(20% r 5 digits)
*
2
*
1
In this range the reading maybe rolling within specification.
*
2
Specify reading < half full scale of range.
Note:
The meter selects the proper range automatically. Each measurement takes about one
second per range, Readings >40.00
μ
F will take
≈
8 seconds or greater
Form Number
2
TM61340-2
Rev
November 2007