SECTION 6 TROUBLESHOOTING
LifeCare PCA
6 - 6
Technical Service Manual
6.3.2
ERROR
CODES
REQUIRING
TECHNICAL
SERVICE
lists infusion system error codes that require technical service. Also listed
in
are malfunction descriptions, possible causes, and corrective actions. Alarms
associated with malfunctions do not allow the audio portion of the alarm to be paused.
Table 6-2. Error Codes Requiring Technical Service
Alarm
Code
Malfunction Description
Possible Cause
Corrective Action
170
Interrupt Overlap
Interrupt failure
171
Invalid Interrupt
Invalid or unused interrupt
172
Stack Error
RAM error
173
Task Dead Error
Periodic task or interrupt
stopped running
174
Kernel Error
Kernel failure
175
Watchdog Error
MCU fails to strobe the
watchdog within its timeout
period
Verify jumper on pins
5 - 6 of JP1 on MCU PWA
If problem persists,
176
Cancel Timer Error
Timer message has been
unsuccessfully cancelled
200
Motor Overdelivery Condition
Motor is running faster
than expected
201
Motor Underdelivery Condition
Motor is running slower
than expected
202
Motor Runaway
Motor is running without
permission
203
Excessive Rate
Rate/Mode Mismatches
Rate is greater than
the maximum rate
204
Mode
Unexpected mode
205
Motor Step Time Error
Motor step time out of range
206
Rate
Unexpected rate
207
Delivery Volume Error
Volume infused out of range
310
5-Volt Error
Two readings outside
range of 4.7 V - 5.3 V
Replace power supply PWA
311
Lithium Battery Error
Lithium coin cell battery
voltage below acceptable
minimum value
Verify lithium coin cell
battery connection
Replace lithium coin cell
battery