SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 32
Technical Service Manual
5.5.9
BARCODE
READER
WINDOW
INSPECTION,
TEST,
AND
CLEANING
Inspect the barcode reader window and test the barcode reader at least once every
12 months
.
The recommended cleaner for this procedure is CleanTex CT 811 OpticPad Optic Surface
Cleaner.
To inspect and clean the barcode reader window, see
, and proceed as follows:
1. Place the infuser on a flat, stable surface.
2. Ensure the barcode reader assembly is aligned with the barcode reader window.
If the barcode reader assembly is loose or misaligned, align it and tighten the screws.
3. Inspect the barcode reader window to verify it is free of damage and there
is no evidence of fluid ingress. If the window appears to be damaged,
replace the barcode reader window
.
4. Confirm that the barcode reader window is clear. If the window is not clear, follow
the cleaning instructions listed below:
• Use recommended CleanTex CT 811 OpticPad Optic Surface Cleaner or a soft,
lint-free cloth moistened with a streak-free optical glass cleaner for cleaning.
• To avoid fluid ingress, do not spray fluids or cleaning solutions on the barcode
reader window, and press lightly on the barcode reader window while
cleaning.
To test the barcode reader, access the
System Tests
in
Biomed Mode
.