APPENDIX
Technical Service Manual
A - 3
LifeCare PCA
ELECTROMAGNETIC
IMMUNITY
FOR
ME EQUIPMENT
AND
ME SYSTEMS
THAT
ARE
NOT
LIFE
‐
SUPPORTING
provides guidance for use of the infusion system near communications
equipment. The infusion system is intended for use in the electromagnetic environment
specified below in
. The customer or user should assure that it is used in such
an environment.
Table A-3.
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity for ME Equipment and ME Systems that are not Life-Supporting
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the infusion
system, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
[V
1
] V
Recommended separation distance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
[E
1
] V/m
Recommended separation distance:
Where
P
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the following symbol
d
3 5
,
V
1
---------
P
=
d
3.5
E
1
-------
P
80 MHz to 800 MHz
=
d
7
E
1
------
P
800 MHz to 2.5 GHz
=