7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 17
LifeCare PCA
Figure 7-6. Front Enclosure Assembly Internal Components
7.3.8.6
DOOR
LATCH
ASSEMBLY
REPLACEMENT
Recommended tools for this procedure are a 7/8 inch open end wrench, medium size
flat blade screwdriver, and 1/4 inch nutdriver.
Replacement parts for this procedure are:
Assembly, Door Latch
Screw, 6-32 x 5/16, Hex Head, with Washer
To replace the door latch assembly, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Disconnect the door lock cable from J12 on the MCU PWA.