SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 8
Technical Service Manual
7.3.4
RUBBER
FOOT
PAD
REPLACEMENT
Recommended tools and materials for this procedure are an X-acto knife and isopropyl
alcohol.
The replacement part for this procedure is:
Pad, Rubber Foot
To replace the rubber foot pad, see
, and proceed as follows:
1. Disconnect the infuser from AC power, and place the device on its side.
2. Using the X-acto knife, remove the rubber foot pad and scrape the enclosure recess
to remove adhesive residue. Each adhesive-backed rubber foot pad is bonded
in its recess. Do not damage the recess.
3. Using isopropyl alcohol, clean any adhesive residue from the enclosure recess.
4. Remove the protective backing from the self-adhesive surface of the replacement
foot pad and press the pad in place.
5. After approximately five minutes, verify the foot pad is secure.
Replacement of a rubber foot pad is routine maintenance and no verification procedure
is normally required. However, if the infuser may have been damaged during this
procedure, perform the PVT in
.
Figure 7-4. Battery Assembly, AC Power Cord, and Patient Pendant Assembly
FRONT ENCLOSURE
#6 LOCK WASHER (3)
6-32 X 1/2
PAN HEAD
SCREW (4)
BATTERY ASSEMBLY
PATIENT PENDANT
ASSEMBLY
6-32 X 1/4 PAN HEAD SCREW
BATTERY
DOOR
BATTERY
DOOR PAD
SPLIT LOCK
WASHER
REAR ENCLOSURE
AC POWER
CORD SET
RUBBER FOOT PAD (4)