26
FOB-BA-e-2124
Models FOB-LM:
•
Following each calibration, the integrated counter value (<Con x>) is
increased by one.
•
If the service menu is brought up after the password has been entered, the
verification is lost, as the value of the integrated counter (<Con x>) is
increased by ‘1’.
Calibration procedure:
•
If the calibration menu is brought up after the password has been
entered, the verification is lost, as the value of the integrated counter
(<Con x>) is increased by ‘1’.
Balance indication
Operation
Switch the balance off.
Press the
ON/OFF
button and hold it down.
When the
ON/OFF
is held down and Con x value is
displayed, press
UNIT
3 times. Then release the
ON/OFF
button.
A "0000" password request will be displayed.
Enter "9810" or “9788”password.
When you press
TARE
, the numerical value will be
increased; when you press
UNIT
, the decimal position
is changed.
To confirm the entered data, press
ON/OFF
button
The integrated counter value will be displayed.
Press
TARE
to confirm, "0
0
" will be displayed.
(example)
Put on a calibration weight (see chapter. 1 Technical
data). Wait until the stabilisation indicator is displayed,
press
TARE
to confirm.
"CAL" will be displayed.
The balance will switch to the weighing mode.
Switch the balance off and then on again.
Con value will be increased by "1".
This means the calibration has been completed
successfully.