Firmware Version 1.2
MultiSystem 4070
123
Special functions
DEU
Coupling of several instruments
You can couple several
MultiSystem 4070
measuring instruments and in-
crease the number of available input channels with almost no limitations. But
please be aware that the parameters scanning rate, recording time and pre-
triggers must be programmed identically at all coupled measuring instruments.
Connecting a measuring instrument electrically
Coupling of two
instruments
Use the connection cable TKZ 8824-F2-00.50 and connect the digital input/
output jacks.
Coupling of several
instruments
Use the connection cable for external trigger TKZ 8824-D8-04.00 and couple
the instruments serial or parallel:
You can also couple
MultiSystem 4070
instruments with
MultiSystem 5060 Plus
instruments.
Hinweis
Malfunctions possible!
Only use the connection cables available from HYDROTECHNIK to connect the
instruments.
Otherwise there is the risk of malfunctions.
A
Digital input/output
A