10.1.1 Filling with the Filling Tube
1. Press down the pump lever as far as it will go (see fig. 7).
2. Sink the filling tube into the sample.
3. Slowly release the pump lever.
Fig. 7: Filling sample with the filling tube
10.1.2 Filling with a Syringe
IMPORTANT:
The instrument comes with the pump lever without lock
function already mounted. For filling with a syringe, replace this pump lever
with the pump lever with lock function (supplied with the instrument) as
described in section 11.6.
1. Mount the syringe adapter (see section 5.2).
2. Lock the pump lever.
Push the pump lever down and forward simultaneously to minimize the
dead space within the pump.
WARNING
If you remove the pump while filling with a syringe, or if you remove the
syringe directly after filling, the system will become leaky. Harmful liquids
leaking from the instrument may cause injuries.
• Do not remove the pump during routine operation.
• Do not remove the syringe during measurement.
1
2
3