Appendix A: Technical Data
24
E15IB001EN-N
Appendix A: Technical Data
A.1: Specifications
DMA 501
DMA 1001
Density
Measuring range
0–3 g/cm
3
Repeatability s.d.
a
a
According to ISO 5725
0.0002 g/cm
3
0.00005 g/cm
3
Reproducibility s.d.
0.0004 g/cm
3
0.00007 g/cm
3
Accuracy
b
b
Under ideal conditions and for low densities/viscosities
0.001 g/cm
3
0.0001 g/cm
3
Temperature
Measuring range
15 °C to 40 °C (59 °F to 104 °F)
15 °C to 60 °C (59 °F to 140 °F)
Repeatability s.d.
0.1 °C (0.2 °F)
0.02 °C (0.04 °F)
Accuracy
0.3 °C (0.6 °F)
0.05 °C (0.09 °F)
Measurement
Operating pressure sample
max. 10 bar (145 psi) absolute
Sample volume
< 2 mL (incl. filling adapter)
Sample viscosity
0–35,000 mPa·s
Viscosity correction
yes, over the full range
U-View™
yes
FillingCheck™
yes
Data memory
5000 results