C500
|
C520 Series Controller
Conductivity
TDS
Resistivity
Salinity
Operating Manual
Page 1: ...C500 C520 Series Controller Conductivity TDS Resistivity Salinity Operating Manual...
Page 2: ...tion 11 Factory Default Current Outputs 12 Calibration 12 Field Calibration 13 Calibration 13 Measurement Settings Quick Start 15 Temperature Calibration 16 Point Calibration 17 Field Calibration 18 R...
Page 3: ...3 The analyzer has been stored for a long time in an environment where the temperature exceeds 70 C The analyzer must be installed by licensed professionals in accordance with relevant local specifica...
Page 4: ...anual Outputs Two 2 4 20mA 20 4mA 0 20mA Communication RS485 MODBUS RTU Relay Control Contact 5A 240VAC 5A 28VDC Power Supply 9 36VDC std 85 265VAC Power Consumption 3W Environmental Temperature 14 14...
Page 5: ...ion Output G TX RX RS232 Communication Output P P DC Power T2 T2 Temperature Connection EC1 EC2 EC3 EC4 Conductivity Resistivity Connection RLY3 RLY2 RLY1 Three Relays L N Communication Terminal L Liv...
Page 6: ...520 Series Operating Manual 6 Dimension Instrument Installation 1 Cable Recommended stripping length for cables is at least 90mm please use 0 5 to 1 square meter wire 2 Cable ties 3 Waterproof cable g...
Page 7: ...C500 C520 Series Operating Manual 7 Slim Line Wall Mount Bracket...
Page 8: ...Measurement Mode Programming Mode Calibration Mode Note New Sensors are Factory Calibrated Data Logging Trend Chart 2023 01 09 12 53 17 Calibration Set points Output History System 25 0 C 83 01 S cm...
Page 9: ...Temperature Temperature Sensor NTC2 252 k NTC10 k Pt100 Pt1000 Default Temperature Offset 0 0000 Enter Value if Required Temperature Input Automatic Manual Temperature Unit C F Default Calibration St...
Page 10: ...esis 00 000 Default can be modified Relay 2 Status ON OFF High Low Setpoint High Alarm Low Alarm Clean Limit Value 00 00 Default can be modified Hysteresis 00 000 Default can be modified Output Curren...
Page 11: ...ight Power Saving Enter Value Long Bright Information Soft Version 19 1 0 Password Settings 0000 Serial Number Factory Default No Keep User Settings Yes Reset back to Factory Default Factory Calibrati...
Page 12: ...en factory calibrated before delivery and users can use it directly Note Calibration if required can be made using at least one of the five calibration points Not all points are required for sensor ca...
Page 13: ...t measurement if there are errors the error data can be modified by this function Linear adjustment Linear values after field calibration are stored in this item with factory data of 1 00 The main val...
Page 14: ...he movement direction of the key Press to switch the displacement direction and press to switch the display range enlarge shrink Item Display Selection Sub Menu System Terminal Current Tuning Current...
Page 15: ...r Back Press ENT key for Save Press ESC key for Back Press ENT key for Save Press ESC key for Back Press ENT key for Save Press ESC key for Back Press ENT key Press ENT key Select Configure Press key...
Page 16: ...T key for Save Press ESC key for Back Press ENT key Select Configure Press key Temperature Press ENT key Select Temperature Press key Press ENT key Select Temp Input Press key Press ENT key Select Tem...
Page 17: ...Select Point 3 Using Press key Press ENT key Select Point 4 Using Press key Press ENT key Select Point 2 Press key 0000 Enter Value Press ENT key for Save Press ESC key for Back Calibrating Select Val...
Page 18: ...ss ENT key Select Point 3 Press key Press ENT key Select Point 2 Press key Std Cal Field Cal Press ENT key Select Temperature Press key 0000 Enter Value Press ENT key for Save Press ESC key for Back C...
Page 19: ...Press ENT key Select Limit Value Press key Press ENT key Select Hysteresis Press key Press ENT key Select High Low Set Press key OFF On Press ENT key for Save Press ESC key for Back Select Item Press...
Page 20: ...Configure Press ENT key Select Output Press key Current 2 Press ENT key Select Current Press key Press ENT key Select Lower Limit Press key Press ENT key Select Upper Limit Press key 0 20mA 20 4mA 4...
Page 21: ...ess ENT key Select Parity Check Press key 9600 BPS 19200 BPS 4800 BPS Press ENT key for Save Press ESC key for Back Select Item Press key Odd Even None Press ENT key for Save Press ESC key for Back Se...
Page 22: ...ey Password Serial Value 000000 Press ENT key for Save Press ESC key for Back Value English Std Low High Med Press ENT key for Save Press ESC key for Back Select Item Press key Display Speed Press ENT...
Page 23: ...ayer Figure 1 MODBUS Protocol Data Unit MODBUS protocol mapping on a specific bus or network introduces additional fields of protocol data units The client that initiates the MODBUS exchange creates t...
Page 24: ...this order from left to right the least significant bit LSB Maximum Significant Bit MSB Start bit 1 2 3 4 5 6 7 8 Stop bit Stop bit Figure 3 RTU pattern bit sequence Check Domain Structure Cyclic Redu...
Page 25: ...3 5 characters 1 247 Function codes conforming to MODBUS specification Data conforming to MODBUS specification CRCL CRCL No signal bytes during 3 5 characters Byte 3 5 1 n 1 1 3 5 Figure 7 MODBUS def...
Page 26: ...Value Low Bytes 110 0x00 Function code 0x64 Figure 10 Examples of read and hold register request and response frames MODBUS function code 0x10 write multiple registers This function code is used to w...
Page 27: ...number in binary form first finding the binary representation of the integer part 17 Decimal 16 1 1 24 0 23 0 22 0 21 1 20 The binary representation of integer part 17 is 10001B then the binary repres...
Page 28: ...10 0111 1011 0110 0110 0110 0110B to decimal number Step 1 Divide the binary float ing point number 0100 0010 0111 1011 0110 0110 0110 0110B into symbol bit exponential bit and mantissa bit 0 10000100...
Page 29: ...0 then shortcircuit the terminal Input and Ref the instrument should be displayed in 7 000 or within the error range If the display is not correct please contact the manufacturer for maintenance Pleas...
Page 30: ...one year from the date of purchase If repair is necessary and has not been the result of abuse or misuse within the warranty period please return to Icon Process Controls and amendment will be made w...
Page 31: ...31 23 0033 01 27 23 Need a great sensor Pair your C500 C520 Series controller with the industry s most innovative and reliable sensing technology iconprocon com...