background image

Page 8 

7. Bias Adjustment  

Please regularly check this important function! 

EASY WAY: 

With the amplifier warmed up and on zero 

volume. Measure the 0.5v DC (500mv) voltage on each 
KT88 test point by using the supplied meter on the 
indicated range. The black lead connected to the “earth” 
or one of the “negative” Speaker terminals as shown. 
Note the approximate readings. If they are all similar in 
the region of 430-500mv this is fine. If one or two are out 
of line, adjust to be similar to the others. Or you could re-
set them all to 490-500mv. This will give the best 
performance. 

If you are unsure about any aspect contact your 
retailer, Icon Audio or a competent service engineer. 
You are welcome to speak to one of our engineers 
by phone or email. 

 

The Stereo 60 uses the ‘Fixed bias’ method of valve 
operation. This has the advantage of higher power, and 
cooler running. However occasionally (once or twice a 
year) it is advisable to check the bias reading using the 
supplied meter to ensure best performance from the 
amplifier. This is a safe procedure which involves 
measuring 500mv (or 0.5v) in the four sockets next to 
each KT88 and adjusting if necessary. 

 

1, Tools you will need: The supplied meter or one set to 
measure approx 500mv, and a small flat blade 
screwdriver. Adjustments are done at zero volume with 
speakers connected. Run the amp for at least 15mins (if 
it is working correctly). 

 

2, Connect: the black probe to the chassis ‘earth’ by 
unscrewing the ‘0’ speaker terminal and tightening the 
probe in the exposed hole. And the other in the test 
socket adjacent to the valve on test. Set the ‘Icon’ meter 
to 2000mv or the ‘black mark’. See pics. 

 

 

Making the ‘earth’ connection. 

3, Checking: You should get a reading of 430-500mv. 
But mains voltage fluctuations can affect your readings 
up to 10%. It is more important that both valves of a pair 
read within about 25mv. 

In some areas fluctuating 

mains may make measurement difficult, in which 
case use best estimate or use an analogue type 
“AVO” meter.

 

4, Adjusting: If not 500mv, set this by using the bias 
adjuster adjacent, then check the other valve of the pair, 
(L to R 1&2, 3&4). The bias adjustment pots from left to 
right are normally ‘mirror image’ in rotation, so some will 
be anti-clockwise and some clockwise. They are very 
sensitive so adjust very carefully; use tiny turns of the 
screw. If the reading appears a little unstable this is 
normally due to mains fluctuations. 

5, If one or more valves are showing very erratic 
readings or you cannot set the maximum to 500mv, 
then that valve is probably faulty or out of 
specification. If you are unable to set the reading high 
enough this means the emission of the valve is low. 

No reading (or 0) may also mean a blown fuse 
see “Valve fuses” under section 5. 

 

 

Showing a probe reading 1

st

 output valve 

 

Replacing the KT88 Valves

 

Icon Audio are happy to replace valves and check to 
performance of your amplifier, and advise on the 
latest upgrades available. If you decide to do this 
yourself please read the following notes first. 

Important: Do not attempt to change the KT88 
without reading these notes. Failure to do so 
could be both dangerous and damaging to the 
amplifier. 

 

 

Health & safety; High voltages are present inside 
the amplifier and on exposed valve sockets when 
valves are removed, so take suitable care. It is 
not necessary to remove the bottom cover. 
Beware valves get hot in operation! 

 

5, Changing valves: You should if possible check the 
bias setting before you attempt to change the 
valve(s), in order to familiarise your self with the 
procedure. The safe way is to change and check one 
or two valves at a time. Remove the first old valve 
and fit the replacement. Switch on and measure bias, 
you should be setting the reading for each valve to 
about 500mv, Do not allow the reading to go above -
900mv. Don’t worry how low the reading goes this 
will not cause damage. Continue in the same way 
and fit all four valves. Do final adjustment when the 
amplifier is fully warmed up. 
If all is well there should be no more than a barely 
detectable hum from the speakers, and the amplifier 
should sound OK When tested. 

 

6, If you cannot set up 500mv then the valve is 
probably faulty or is unsuitable.  
If the valves are brand new, you will need to check 
again after approximately 10 & 100 hours, after that 
only occasionally or if you suspect a problem.  

 

7, To avoid damage to the amplifier and electric 
shock hazard you must use only valves marked 
KT88 (or direct equivalent e.g. 6550). KT66s may 
also be used with 50w output power, 6SL7 6SN7 
or that you know to be direct equivalents.

 Use 

only valves which you know to be new or good 
condition and test the amplifier thoroughly before 
resuming normal use.  

 

Summary of Contents for Stereo 60 MK 2

Page 1: ...wiring hand wiring without the use of printed circuit board s results in an open euphoric sound that is wonderfully detailed and warm sounding In order to get the best out of your amplifier please rea...

Page 2: ...Output Valve Bias level 500mv Sound Quality Valve Microphony Valve Seating Hum level left right mv mv RF Test Power out 1 kHz 8ohms L R Channel Balance LED brightness Serial No sticker and recorded M...

Page 3: ...the switch should be DOWN in the low sensitivity position We would recommend use of the tape input with the volume on FULL or set as required and volume adjusted from your preamplifier although if yo...

Page 4: ...00v DC which when switched off can take several minutes to discharge Do not remove bottom panel unless you are a competent engineer There are no user serviceable parts inside Like other household elec...

Page 5: ...better and be cheaper to re arrange your room and use shorter cables than to spend a fortune on longer cables You can either hard wire your cable to the amplifier by baring enough cable to fit in the...

Page 6: ...Valves Good quality valves should sound better and have a good service life maintain their performance and be reliable the last three items will make a valve sound better longer The upgrade valves su...

Page 7: ...ne us You could also return the valves to Icon UK for testing free of charge Carefully remove the valves the KT88s should be held by the base when removing to prevent damage numbering them with a mark...

Page 8: ...ockwise They are very sensitive so adjust very carefully use tiny turns of the screw If the reading appears a little unstable this is normally due to mains fluctuations 5 If one or more valves are sho...

Page 9: ...es are classed as consumables on a 12 month reducing percentage of the cost of replacement LEDs used for aesthetic and d cor purposes are not covered These may be disconnected if not required This amp...

Page 10: ...edure 1 With the top packing in place turn the box upside down with the flaps open 2 Lift the cardboard box off 3 Remove the bottom packing first 4 Turn the unit the right way up and remove the top pa...

Reviews: