icon Audio Stereo 300 MKII Instruction Manual And Users Manual Download Page 6

 

Page 6 

5 Trouble Shooting 

1. Amplifier Dead 

Check  the  power  fuse  at  the  back  of  the  amplifier.  To 
gain  access,  remove  the  mains  lead.  The  fuse  is  in  a 
small  plastic  drawer,  which  forms  part  of  the  socket 
assembly.  To  open  insert  a  flat  bade  screwdriver  or 
similar  and  prise  open. 

The  fuse  in  use  is  the 

innermost

 the outer is a spare. Should the replacement 

fuse also blow there is a fault contact you dealer or Icon 
Audio.  Replace  with  the  same  type  Available  from  Icon 
Audio  free  in  the  UK  (outside  the  UK  you  must  pay 
postage).  
The  fuse  in  the  power  cord  plug  (if  any)  should  be  a  5 
amp  fuse,  although  unlikely,  this  should  be  checked  if 
the amplifier fuse is OK. 

2. No sound 

Have  you  selected  the  right  input?  Are  the  connections 
OK?  Is  everything  switched  on?  Are  the  speakers 
connected? 

Is the amp in “standby”? 

If you think you have tried everything and the meter light 
is  working  and  the  heaters  are  glowing  either  or  one  of 
the  internal  fuses  may  have  blown.  They  are  located 
inside  the  amplifier  and  should  be  replaced  by  a 
competent engineer. Replacements free on request from 
Icon  Audio.  Contact  us  or  your  dealer  for  further 
guidance.  If  a  second  fuse  blows  the  amplifier  needs 
service. 

Distorted sound. 

One  channel  or  both?  Try  another  source;  if  sound 
improves  then  it’s  probably  something  wrong  with  the 
first  source.  If  no  improvement  try  different  speakers,  if 
no  improvement  could  be  an  amplifier  problem.  If  one 
channel  try  swapping  the  input  cables  over  if  the 
distortion does not move could be an amplifier fault. If so 
double  check  by  swapping  the  speaker  cables  over  If 
fault  moves  to  the  other  speaker  this  will  confirm  an 
amplifier fault. It fault  does not move this would indicate 
a speaker problem. 

Hum Problems and Changing Valves.  

The 300B filaments are heated by AC. This causes hum 
to  be  introduced  into  the  circuit  so  with  “push-pull”  the 
hum  of  one  valve  is  adjusted  to balance  out  the  hum  of 
the  other  valve  of  the  push-pull  pair.  This  does  not 
normally  require  adjustment  but  when  the  valves  are 
changed  you  may  notice  an  increase  in  low  level  hum, 
requiring the following adjustment to be made. If a meter 
which  reads  AC  at  1-10mV  is  available  connect  to  the 
loudspeaker  terminals  with  the  loudspeaker  still 
connected  and  adjust  for  minimum  hum.  Normally  this 
would be 1-2mV. 
Alternatively, this could be done with one person making 
the adjustment and another listening  closely to the bass 
speaker  for  minimum  hum.  The  volume  should  be  at 
“zero”. Only a small adjustment should be necessary. 

Other Causes of hum.

  

1.  Establish  if  the  hum  is  internal  or  external  to  the 
amplifier. If the hum disappears with no inputs connected 
the problem is with the source and/or interconnects. 
2.  If  the  amplifier  has  serious  hum,  this  may  be  a  fault. 
Contact your dealer or Icon Audio. 
3.  Hum  Loop.  If  the  hum  is  only  on  one  input,  you  may 
have  a  “hum  loop”  caused  by  too  many  “ground”  or 
“earth” connections. A possible remedy for this is to use 
an interconnect which only has the screen connected at 
one  end.  Other  causes  of  low-level  hum  can  be  from 
adjacent  equipment,  so  experiment  with  moving 

equipment around to see if this makes the hum better 
or worse. 

Interference Problems 

This  amplifier  design  is  very  resistant  to  mains-born 
interference.  But  some  heating  systems,  fridges, 
cookers  etc.  can  generate  RF  (radio  frequency) 
interference which travels through the air (and walls). 
Although  rare  this  can  be  very  irritating.  Relocating 
the equipment will often help. Or a simple suppressor 
capacitor  will  often  cure  this.  Mobile  phone 
interference is normal for up to 1-2 meters away. 

One channel missing. 

Usually  ‘bad’  connection  on  either  the  input  or  the 
speakers.  Try  swapping  the  connection  over  to 
establish if the problem is caused by: 
(a) An input to the amplifier: 
 

Sound will move to the other channel. 

(b) The amplifier or speakers. 
 

Sound will not move. 

(c) Consider checking internal fuses (see 5.2). 

Strange noises coming from speakers: 
Turn volume to minimum on unused input, if problem 
corrected  either  fault  with  source  unit  or  with 
connection. If noise persists, problem with amplifier. 
If  a  whole  output  valve  glows  red  (other  than  the 
heater),  often  accompanied  by  a  hum  through  the 
speakers,  switch  off  immediately,  and  refer  to  Icon 
Audio  or  a  service  engineer,  as  this  could  be  valve 
failure. A valve that is hot is not a guarantee that it is 
working  properly;  conversely  a  valve  that  is  not  hot 
will not be working (usually cold to the touch). 
“Rustling”  and  “crackling”  noise  is  often  due  to  the 
smaller valves, which if at low level if at a low level is 
nothing to nothing to worry about. 

 

Valve Life and Replacement 

 

Valve  life  will  depend  upon  such  things  as  hours  of 
use  and  number  of  on/off  cycles,  Do  not  switch  on 
and  off  unnecessarily  (see 

Standby).

  The  300B 

valves  are  operating  in  “CLASS  A/B”  mode  (approx. 
5  Watts  output)  which  means  they  are  using 
moderately high current all of the time. Therefore we 
would suggest for use of  more than 3  hours per day 
replace the valves every 5 years.  
If  using  expensive  300Bs  the  valve  life  may  be 
extended by running the 300Bs at a lower HT to give 
around  20  Watts  output.  Icon  Audio  or  most 
engineers will be able to do this. 

 

Service: 

Should  you  suspect  a  problem,  you  could 

return the unit to Icon Audio for a periodic service or 
return the valves for testing free of charge.  Carefully 
remove the valves, hold by the base when removing, 
to prevent damage Make sure they are numbered so 
they  may  be  replaced  in  the  same  position.  They 
should  be  well  packed  in  cardboard  &  foam  or 
similar,  and  returned  to  Icon  Audio  for  testing. 
(Valves are very rugged if packed properly). 

 

Mains Supply 

This  amplifier  is  hard  wired  to work  on  230/240v  AC 
(or  117v  US).  The  transformer  may  easily  be  re-
configured  for  110/120v  ac.  Contact  for  more 
information. 

 

 

Summary of Contents for Stereo 300 MKII

Page 1: ...tages employing the venerable 6SN7 CV181 giving an excellent combination of the best of vintage valve sound yet accurate and dynamic to suit modern speakers and recordings In order to get the best out...

Page 2: ...ernal wiring check Check Triode mode N A Run 24 hour test Check inputs tape monitor Output Valve Bias level mv Sound Quality Channel Balance Valve Microphony Valve Seating Hum level left right mv RF T...

Page 3: ...ndby switch into the down position The amplifier should now be working 300Bs do not glow you can only see this by looking down vertically inside the valve With the volume control set to minimum fully...

Page 4: ...this is personal taste Icon or your dealer will advise you As all speaker cables have losses keeping the speaker cables shorter than 10m will help You can either hard wire your cable to the amplifier...

Page 5: ...eft in the standby mode this will save power Do not leave the amplifier switched on 24 7 Your valves will be worn out in approximately nine months Never leave the amplifier switched on when unattended...

Page 6: ...dio 3 Hum Loop If the hum is only on one input you may have a hum loop caused by too many ground or earth connections A possible remedy for this is to use an interconnect which only has the screen con...

Page 7: ...nough this usually means the emission of the valve is too low Replacing the 300B Valves Important Do not attempt to change the 300Bs without reading these notes Failure to do so could be both dangerou...

Page 8: ...caps Internal wiring using PTFE silver plated cable Valves carefully matched for best performance Gold plated Input speaker terminals Inputs for CD Tape Tuner Aux iPhone Record loop with monitor swit...

Page 9: ...ioning correctly it may be helpful to contact us first as we may be able to assist you Then we would request that you return the item to us for further action You are advised to inform us of any chang...

Page 10: ...n 2 Lift the cardboard box off 3 Remove the bottom packing first 4 Turn the unit the right way up and remove the top packing Please be careful not to damage the small switches on the front panel When...

Reviews: