4 - 13
ADJUSTMENT ITEM
ADJUSTMENT CONDITION
OPERATION
VALUE
[V/U ATT20dB BPF7 (H) f]
• Set the SG as;
Frequency
: 499.9815 MHz
Push [F-7] key.
(Automatic
adjustment)
[V/U ATT20dB BPF8 (L) f]
• Set the SG as;
Frequency
: 500.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF8 (M1) f]
• Set the SG as;
Frequency
: 562.5215 MHz
Push [F-7] key.
[V/U ATT20dB BPF8 (M2) f]
• Set the SG as;
Frequency
: 625.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF8 (M3) f]
• Set the SG as;
Frequency
: 687.5215 MHz
Push [F-7] key.
[V/U ATT20dB BPF8 (H) f]
• Set the SG as;
Frequency
: 749.9815 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (L) f]
• Set the SG as;
Frequency
: 750.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (M1) f]
• Set the SG as;
Frequency
: 820.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (M2) f]
• Set the SG as;
Frequency
: 910.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (M3) f]
• Set the SG as;
Frequency
: 990.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (M4) f]
• Set the SG as;
Frequency
: 1070.0215 MHz
Push [F-7] key.
[V/U ATT20dB BPF9 (H) f]
• Set the SG as;
Frequency
: 1149.9815 MHz
Push [F-7] key.
NOTE:
HOLD