8000 SERIES IR CAMERA USER MANUAL

14

L.

Plug the Mini-USB end of the USB cable 

into the back of the 8000 Series Camera.

M.

N.

Plug the other end of the USB 

cable into the laptop computer.

Power on devices.

ALLOW  AT  LEAST  45  MINUTES  FOR  THE  TEMPERATURE 

REFERENCE SOURCE TO WARM UP. THIS WILL PROVIDE THE 

MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS. 

SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS 

PURCHASED FROM ICI AS COMPLETE SYSTEMS AND IT WILL 

LAUNCH AUTOMATICALLY ON STARTUP.  AFTER THE SOFTWARE 

LAUNCHES THE USER SHOULD ENSURE THE TARGET ZONE IS 

IN VIEW OF THE CAMERA AND THE TEMPERATURE REFERENCE 

SOURCE IS POSITIONED OFF CENTER TO KEEP THE PERSON BEING 

IMAGED AS CENTERED AS POSSIBLE.

Summary of Contents for 8000 Series

Page 1: ...t Control THIS DEVICE IS INTENDED FOR ADJUNCTIVE USE WITH OTHER CLINICAL DIAGNOSTIC PROCEDURES TO MEASURE HUMAN BODY TEMPERATURE VIA NON CONTACT SKIN MEASUREMENTS VISUALIZED FROM THE HUMAN FACE NOT MEANT FOR STANDALONE CLINICAL DIAGNOSTIC PROCEDURES OR TO TREAT OR DIAGNOSE PATIENTS For more information contact us 409 861 0788 sales infraredcameras com www infraredcameras com Medical Medical Edit T...

Page 2: ... BUT ARE PRESENTED WITHOUT WARRANTY OF ANY KIND NOTWITHSTANDING ANY OTHER WARRANTY HEREIN ALL DOCUMENT FILES AND SOFTWARE ARE PROVIDED AS IS WITH ALL FAULTS ICI DISCLAIMS ALL WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITHOUT LIMITATION THOSE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT OR ARISING FROM A COURSE OF DEALING USAGE OR TRADE PRACTICE THE SOFTWARE LICENSE AND L...

Page 3: ... Information 3 Intended Use 4 Technical Specifications 5 Structure 6 Package Contents 7 Installation Instructions 8 Computer Setup 9 Cleaning and Maintenance 10 Site and System Validation Checklist 11 Troubleshooting 12 About ICI Contents 4 4 4 5 7 8 9 17 18 19 21 23 ...

Page 4: ...re during use storage and transportation to prevent dangerous actions such as the device being hit by external force or falling from heights During transportation and storage the original packaging box must be used Prior to start of the device make sure that the power supply is properly connected If the power supply is connected incorrectly the device may be damaged Do not place any objects on the...

Page 5: ... 25 C to 45 C 77 F to 113 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 80 C 40 F to 176 F Humidity 5 to 95 non condensing Accuracy 0 2 C 0 36 F Pixel Operability 99 Shock Vibration 75 G 4 3 G Dimensions without lens 44 5 mm x 44 5 mm x 46 5 mm L x W x D 0 5 mm 1 75 x 1 75 x 1 83 L x W x H 0 02 Power 5 V DC 500 mA 1 W 5 V DC 500 mA 1 2 W Weight without lens 72 g 2 5 oz Interfac...

Page 6: ... 77 F to 113 F Operation Range 0 C to 50 C 32 F to 122 F Storage Range 40 C to 80 C 40 F to 176 F Humidity 5 to 95 non condensing Accuracy 0 2 C 0 36 F Pixel Operability 99 Shock Vibration 75 G 4 3 G Dimensions without lens 44 5 mm x 44 5 mm x 46 5 mm L x W x D 0 5 mm 1 75 x 1 75 x 1 83 L x W x H 0 02 Power 5 V DC 500 mA 1 W 5 V DC 500 mA 1 2 W Weight without lens 72 g Interface USB Mini B USB 2 0...

Page 7: ...8000 SERIES IR CAMERA USER MANUAL 7 1 Mini USB interface 1 5 Structure 5 1 Appearance 5 2 Definitions of Housing Interface ...

Page 8: ...ENSURE ALL SYSTEM EQUIPMENT AND COMPONENT ITEMS ARE PRESENT BEFORE BEGINNING INSTALLATION Mounting Bracket Laptop Laptop AC Power Cord Software USB Drive Temperature Reference AC Power Cord Laptop AC Power Adapter 8000 Series IR Camera Mini USB to USB Cable ...

Page 9: ...HE MOUNTING BRACKET 7 Installation Instructions SELECT AN APPROPRIATE AREA FREE OF IMMEDIATE AIRFLOW FROM DOORWAYS AND AIR CONDITIONING VENTILATION SYSTEMS THE SELECTED AREA SHOULD HAVE A STABLE AMBIENT TEMPERATURE BETWEEN 20 C AND 24 C 68 F TO 75 2 F AND RELATIVE HUMIDITY RANGE FROM 10 TO 50 7 1 Tripod Setup Place mounting bracket securely on back of laptop computer ...

Page 10: ...A USER MANUAL 10 D C Tighten mounting bracket by turning the dial to the right If you did not purchase tripod accessories place the laptop on a stable desk or table Mount the mounting bracket to a tripod using the 20 mount ...

Page 11: ...8000 SERIES IR CAMERA USER MANUAL 11 E F G Plug laptop power cord into laptop power adapter Plug laptop power adapter into laptop computer Plug power cord into a 110 120V electrical outlet ...

Page 12: ... tripod using the 20 mount H I Plug power cord into Temperature Reference MAKE SURE TRIPODS DO NOT BLOCK THE DIRECT PATH OF PERSON S TO BE IMAGED TO ENSURE THE EQUIPMENT WILL NOT BE MOVED OR KNOCKED DOWN USING A DIVIDING BARRIER WILL HELP KEEP TRIPODS SEPARATE FROM THE PATH ...

Page 13: ...IR Camera to a tripod mount using the 20 mount J Plug power cord into a 110 120V electrical outlet ENSURE THE CAMERA LENS AND TEMPERATURE REFERENCE SOURCE ARE FACING EACH OTHER ON PARALLEL PLANES MAKE SURE CAMERA TARGET AREA IS FACING A NON REFLECTIVE BACKGROUND ...

Page 14: ...TURE REFERENCE SOURCE TO WARM UP THIS WILL PROVIDE THE MOST ACCURATE SKIN TEMPERATURE MEASUREMENTS SOFTWARE ARRIVES INSTALLED ON COMPUTERS AND TABLETS PURCHASED FROM ICI AS COMPLETE SYSTEMS AND IT WILL LAUNCH AUTOMATICALLY ON STARTUP AFTER THE SOFTWARE LAUNCHES THE USER SHOULD ENSURE THE TARGET ZONE IS IN VIEW OF THE CAMERA AND THE TEMPERATURE REFERENCE SOURCE IS POSITIONED OFF CENTER TO KEEP THE ...

Page 15: ...o laptop 7 2 Charging Instructions C Plug power cord into a 110 120V electrical outlet THE DEVICE TAKES BETWEEN TWO AND THREE HOURS TO CHARGE COMPLETELY IT MAY CHARGE IN LESS TIME DEPENDING ON HOW LOW ITS BATTERY HAS DRAINED YOU CAN USE YOUR SYSTEM WHILE IT S CHARGING BUT THE CHARGING TIME WILL TAKE LONGER ...

Page 16: ...with secondary evaluation methods e g an NCIT or clinical grade contact thermometer Signage should instruct for the removal of glasses headwear and masks Read the current government guidance regarding the use of telethermographic systems for skin temperature measurements Additional information can be found by reading IEC 80601 2 59 2017 Medical electrical equipment Part 2 59 Particular requirement...

Page 17: ...necting the system s computer to the Internet or your secure internal network be sure to have your IT department install your organization s anti virus software anti malware and if applicable system access software and apply security updates as necessary If connecting the system s computer to the Internet be sure to locate the computer behind a firewall Install security updates to the computer s o...

Page 18: ...iddle to the edge paying attention to not crack the lens or use too much liquid If the lens is still not clean replace the cloth and repeat the wiping process 9 Cleaning and Maintenance 9 2 Disinfecting the Camera Surface Do not use corrosive cleaning solutions on the optical glass components It is recommended to disinfect the camera surface regularly with a non corrosive sanitizing product Follow...

Page 19: ...humidity is within 10 50 10 2 The Screening Background Placement is parallel and perpendicular to screening camera At a distance of 12 ft Consists of a non IR reflective homogeneous background Minimum size is 10 x 8 White to gray flat finish 10 3 The Traffic Flow Establish a guided pathway from the entrance to the screening area Design this pathway so there is an equilibration time No restrooms on...

Page 20: ...t up the software according to the instructions Documentation can be found online on the IR Flash STM web page under the Downloads section or at this address https infraredcameras com ir flash stm manual 10 7 Final Step Final validation is dependent upon a signed copy of this checklist from each person responsible for each site Upon completion please sign below ICI representative Company represent...

Page 21: ...software Check that software settings are correct Fixed temperature of 35 C or 37 C 95 F or 98 6 F depending on temperature reference source Temperature reference source has crosshair over it Remove masks glasses Camera is proper distance from person Camera is at proper height Temperature reference source in view perpendicular to camera but not blocked Person is looking directly at camera lens not...

Page 22: ...ER MANUAL 22 11 3 No thermal image Take off the lens cover Ensure device is receiving power 11 4 Thermal imager shuts down or device will not start Charge the device Ensure device is within the safe operating temperature range ...

Page 23: ...nstitute In addition to providing custom germanium silica and sapphire optics we also build windows for enclosures as well as custom pan and tilt units We can even provide customizable explosion proof systems Our knowledge and experience stems from years of using infrared imaging and temperature measurement instruments to provide solutions to managers engineers scientists inspectors and operators ...