IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
52
IceSense3™
1.
Dewar refilling and replacement
2.
Maintain Sterility (Use of sterile sleeves)
3.
Cryoprobe registration and connection
4.
System tests
To prepare the system for treatment, choose the option “Prepare for Treatment” in the Main
Menu as shown in the below figure.
Figure 40: Preparing the system for treatment by pressing “Prepare for Treatment” (green arrow) on Main Menu
screen
5.2.7.1
Dewar
Prepare the Dewar for treatment as per the following instructions and as detailed on the system
screen (see Fig. 40).
Filling the Dewar
Open the Dewar compartment door and take out the Dewar. Carefully cover it with the sponge lid.
Caution
Follow the detailed instructions on open the Dewar storage- when you want to replace a Dewar.
The Dewar must be completely filled prior to each procedure or a refill and is never stored
with liquid nitrogen inside the system at the
end
of the working day.
For proper handling in order to ensure safety, please follow the cryogen supplier instructions.