IceCure FAP 7100000 Instructions For Use Download Page 2

 

 

 

Page 

2

 of 

3

 

 

 

DSR7000000-REV G -IFU FOR CRYOPROBE 

 USA

 

 

IceSense3®, IceCure®, IceSense® and ProSense® are Trademarks of IceCure Medical. 
Ltd. In the United States and other countries. IceCure Medical Ltd. and other parties may 
also have trademark rights in other terms used herein. 

Figure 3:

 

Cryoprobe selection 

 
 

 

 

 

 

Figure 7: For illustration only Color Tag 
labels for cryoprobe and the matching 
introducer 

 
The  cryoprobe  serial  number  (S/N)  appears  on  the 
screen/cryoprobe package/cryoprobe plastic grip.  

 

 

Cryoprobe connection 

 

Never reuse a single-use cryoprobe.

 

 
Connect  the  cryoprobe  to  the  cryohandle,  while  maintaining 
sterility of the cryoprobe: 
1- Remove the plug that covers the cryoprobe connection point. 
2-  Insert  the  cryoprobe  into  the  insertion  point  in  the  handle  as 

shown  on  the  system  screen  and  screw  it  until  a  “NEXT”  button 

appears  on  screen  then  confirm  screwing  by  an  additional  slight 
rotation to confirm that cryoprobe connection is secured. 
3- Remove the cryoprobe tip protector. 

 
 

When done, press “

NEXT

” on the 

screen.  

 

: Cryoprobe connection screen

5

 

ure

Fig

 

 
 

 

Cryoprobe operation 

Perform a functional 

pre-test

 to ensure system efficacy and safety 

as explained in the User Manual. 
If a functional problem occurs or there is any unusual appearance 
(such as frost on the plastic cover near the cryohandle, bubbles or 
any  unusual  appearance),  press  "CANCEL"  on  the  system  screen 
and  follow  system  instructions  until  you  are  required  to  safely 
remove the cryoprobe from the cryohandle. Before operating the 
cryoablation  system,  make  sure  you  have  completed  all  pre-
operational stages. 

 You must NOT allow the freeze process to start before 

the cryoprobe tip is actually within the target tissue.  

 
Before  activating  the  freeze  cycle,  insert  the  cryoprobe  into  the 
target tissue under imaging guidance following the steps: 

1.

 

Confirm longest dimension of the target tissue 

2.

 

Plan  the  trajectory  of  the  cryoprobe  prior  to  placement.  
When clinically safe, the center of the cool zone shall be along 
the longest dimension of the target tissue. 

3.

 

The black safety mark should be completely in the tissue. 

4.

 

In percutaneous approach, perform a 3 mm skin incision (for 
example using #11 blade) before the cryoprobe is inserted. 

5.

 

Position the tip of the cryoprobe at the distal end of the long 
axis of the target tissue, when clinically safe 

6.

 

Center of the cool zone, should be in the center of the tissue 
to be ablated (see figure 2 for the distance of the cool zone 
from the tip of the cryoprobe that is visible under Ultrasound 
or CT imaging) 

7.

 

Maintain sterility and patient safety.   

Be aware of the markings on the cryoprobe: the 

wide black mark

 

closest to the tip is the safety mark. In percutaneous procedures 

it 

must be completely

 

inside the tissue 

to avoid skin burns. The rest 

of  the  marks  indicate  depth  of  cryoprobe  insertion:  each  mark 
equals one centimeter with distinctive markings at 5 and 10 cm 

(50 

& 100 mm). 

  

Figure 6: illustration of FAP7200000 cryoprobe markings on the cryoprobe 

Once you have verified the cryoprobe is located in its right place, 
you may begin 

freezing.

 

Portions of the cryoprobe other than the freeze zone, 
including  the  plastic  cover  that  is  located  near  the 
cryoprobe handle, may become cold and cause tissue 

damage.  If  unwanted  freezing  occurs,  immediately  stop  the 
freezing  process.  In  case  of  frost  on  shaft,  start  Extraction  if 
possible.  If  not,  wait  for  passive  Thaw.  In  both  cases,  use  skin 
protection techniques. 
To prevent injury, cryoprobes must be closely observed during use 
for signs of unwanted freezing.

  

During 

Thaw

, the ice ball melts partially or totally depending on the 

thaw time and the tissue properties. 
 Keep the cryoprobe location steady in the target tissue during all 
of  the  thaw  period.  Control  the  process  under  Ultrasound  or  CT 
imaging system. 
The 

Extraction

  step  occurs  at  the  end  of  every  treatment.  Its 

purpose is to allow the cryoprobe’s removal from the target tissue 

in the fastest and safest way.  
At the end of the extraction step, a message will be displayed on 
system

 

screen. Wait for the message, then gently remove the 

cryoprobe from the target tissue.   
Do not force removal of the cryoprobe from the tissue as it might 
increase the risk of hematoma.  
If the cryoprobe cannot easily be extracted from the tissue, press 

the “

Extraction

” icon on screen to initiate another 

Extraction cycle. 

In case the Extraction 

process isn’t available, wait for passive Thaw.

 

 

 

Before  removing  the  cryoprobe  from  the  tissue,  make 
sure  the  freeze  effect  has  been  deactivated  and  the 

cryoprobe can  be easily withdrawn. Never use excessive force to 
extract the cryoprobe. 

 

Cryoprobe disassembling

 

After  removing  the  cryoprobe  from  the  target  tissue,  and  only  if 
system screen displays a message that it is safe to disengage the 
cryoprobe, detach the cryoprobe from the cryohandle as follows: 

1.

 

Unscrew  the  used  cryoprobe  from  the  cryohandle  and 
dispose of it appropriately. 

2.

 

Remove the single-use sterile cover from the cryohandle. 

3.

 

Close the cryohandle with the covering plug. 

 

Following  each  cryosurgical  procedure,  discard  the  single  use 
devices  (single-use  cryoprobe,  single-use  introducer,  single-use 

Figure 4: The cryoprobe

 

Reviews: