Icare PRO tonometer Quick Guide

ENGLISH

TURNING THE TONOMETER ON AND LOADING THE PROBE

1.

 Press the main button. 

2.

 Go to Measure and press the main button. 

3.

 Partially open the probe package. 

4.

 Insert the probe into the tonometer 

  from partially opened package  

  without touching the probe (4a).  

  Lightly press (4b) the probe into the  

  probe base to lock the probe.  

  Be careful not to bend the probe. 

5.

 Press the main button once to  

  activate the inserted probe. 

MEASUREMENT

1.

 Select the eye you want to measure using the left and right navigation buttons.

2.

 Bring the tonometer near the patient’s eye. **Note: the angle must be 90° in both horizontal and 

  vertical measurement. **

3.

 The distance from the tip of the probe to the cornea must be 3-7mm (1/8 - 9/32 inch). If necessary, 

  adjust the distance using the forehead support.

4.

 Press the main button lightly to perform one individual measurement, taking care not to shake the 

  tonometer. The tip of the probe should make contact with the central cornea.

5.

 Make six single measurements. Once the six measurements are complete, the final IOP is displayed. 

  Save the measurement result by pressing the main button after results!

ONE SHORT BEEP

Indicates a correct single measurement. 

DOUBLE BEEP

Indicates an incorrect single measurement. (See error messages from the instruction manual).

LONG BEEP

Indicates that six succesful measurements are done and IOP result is available on the display.

4a.

4b.

3.

From the main menu, select “Turn off” and confirm by pressing the main button twice.

If not used for three minutes, the tonometer will automatically switch off.

1.

5.

3–7 mm

(1/8 - 9/32’’)

90

˚

Correct measuring positionings with Icare  

PRO are horizontal and vertical.