Icare HOME Patient Guide

ENGLISH

25

www.icaretonometer.com

24

CLINICAL PERFORMANCE DATA

• 

No complications (including corneal abrasions)

were recorded.

• 

The eye pressure ranges measured were ≤16 

mmHg, >16 to <23 mmHg and ≥23 mmHg (mmHg 

is the unit of measure for eye pressure).

• 

The average difference of Icare HOME versus 

the gold standard measurement method 

(Goldmann measurement device that is used in 

eye clinics) was approximately 0.5 mmHg (the 

average pressure measurement in the study was 

approximately 18 mmHg with a range of 5 to 50 

mmHg).

•  Compared to the Goldmann tonometer, less than 

5% of measurements fell outside ± 5 mmHg at 

each pressure range and less than 1% fell outside 

± 7.5 mmHg at each pressure range.

• 

The Icare HOME’s variability (difference of repeat 

measurements) on the same eye was ~10% for 

the low and medium pressure range and ~7.5% 

for the high pressure range.

• 

Only certified users were able to proceed to 

self-testing. 10.7% of the patients failed to pass 

training/certification.

• 

In this study, self-testing was limited to a single 

session 10 minutes after certification.

A study was conducted with 383 patients 

 performing self-measurements with the 

 Icare HOME tonometer. A summary of the 

 study findings are below:

LIFETIME

The expected service life of the device is 5 years. 

Following the instructions described in this 

manual is required during the the expected service 

life. The shelf life of the probes in their intact 

original packaging is 3 years. Please check the 

shelf life (expiration date) from the probe label. 

We recommend that you inspect the device for 

mechanical and functional damage and the safety 

labels for legibility annually/every 12 months. 

The life time of the batteries is about 5000 

measurements.

TECHNICAL DATA

Type:

 TA022

Dimensions:

 approximately 11cm x 8cm x 3cm 

Weight:

 approximately 150g

Power supply:

 2 x CR123 non-rechargeable batteries 

(make sure you use batteries with built-in PTC 

protection, Energizer, Varta, Panasonic or Toshiba 

Lithium 3V CR123A)

Measurement range:

 5-50 mmHg

Accuracy

 (95% tolerance interval relatively to 

manometry): ±1,2mmHg (≤20 mmHg) and ±2.2 

mmHg (>20 mmHg)

Repeatability

 (coefficient of variation): < 8% 

The serial number is located on the inside of the 

battery compartment cover. The lot number of the 

probes is on the side of the probe box and the blister 

packing. There are no electrical connections from 

the tonometer to the patient. All the parts of the 

tonometer are applied parts and the tonometer has 

BF-type electric shock protection. 

Summary of Contents for TA022

Page 1: ...1 www icaretonometer com TA022 035 USA 1 4 Belongs to PATIENT GUIDE ENGLISH...

Page 2: ...he device is stand alone is not networked and does not contain operating system software WARNING Do not change the batteries or probe base when the USB cable is connected WARNING Do not connect the US...

Page 3: ...meter or to report unexpected operation or events contact the manufacturer or the manufacturer s representative WARNING Use of this equipment adjacent to or stacked with other equipment should be avoi...

Page 4: ...For reporting a serious incident with the Icare HOME tonometer or software used with it contact Icare technical service at www icaretonometer com GLOSSARY Goldmann Applanation Tonometry the standard e...

Page 5: ...Very thick or very thin corneas Patients for whom there is already difficulty inobtaining clinical IOP measurements for example due to eyelid squeezing or tremor CONTRAINDICATIONS You should not use...

Page 6: ...SURE indicator REPEAT indicator DONE indicator SERVICE indicator BATTERY indicator Power button Cheek support position indicator Type label Cheek support adjustment wheel Silicone lid USB cover Batter...

Page 7: ...and certified probes made by the manufacturer Do not touch the probe The shelf life of the probes is 3 years Use a new unused probe after measurements for both eyes have been taken You should thorough...

Page 8: ...Make sure the probe base light is green If the probe base light is red make sure you are facing straight ahead i e head held at a 90 angle and tilt the tonometer until the probe base light turns green...

Page 9: ...e light illuminates If your doctor instructed you to measure both eyes repeat steps 1 through 9 for your other eye The probes are single use Use a new unused probe when you do measurements the next ti...

Page 10: ...be base is dirty or damaged BATTERY light is flashing Cause Battery power is low ERROR MESSAGE TROUBLESHOOTING 1 Make sure you are facing straightahead i e head held at a 90 angle 2 Adjust the tonomet...

Page 11: ...one lid USB port Adjustment wheel PRECAUTION Do not use rechargeble batteries PRECAUTION Change the probe base every six months PRECAUTION Do not screw the collar in too tightly to prevent the collar...

Page 12: ...meter in water or any other liquid Start Icare software in your mobile device or PC Connect the tonometer to the mobile device or PC using the USB cable The LOAD and MEASURE lights will flash If no li...

Page 13: ...nts with the Icare HOME tonometer A summary of the study findings are below LIFETIME The expected service life of the device is 5 years Following the instructions described in this manual is required...

Page 14: ...ts where noise is so high that the user cannot hear the error signals Mode of operation continuous SYMBOLS Caution See operating instructions for more information BF type device Single use disposable...

Page 15: ...is below standard requirement batteries are not rechargeable Voltage fluctuations flickering emissions IEC 61000 3 3 Not applicable Icare HOME TA022 batteries are not rechargeable Guidance and manufac...

Page 16: ...than 30 cm 12inches to any part of the device1 including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Interference may occur in the vici...

Page 17: ...ombined with ISO 13485 certified quality system has made us a respected player in our field of expertise The advanced Icare product line offers reliable high precision reproducible accuracy in measuri...

Reviews: