17
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber. Failure to
comply may cause damage to the chamber and/or other components.
Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the surface.
Door adjustment
Under normal circumstances, the chamber door does not require
adjustments. However, if the seal fails (resulting in steam leaking from the
front of the chamber), you may adjust it.
Open the door.
Insert the spanner tool in the gap beneath the plastic cover; use the spanner
to grip the adjusting nut. Turn the nut counter clockwise as the figure below.
This will tighten the sealing plate.
Turn the nut until the sealing plate is tight. If the door knob is too tight, you may also turn
the nut clockwise to loosen it.
Caution: Never adjust the chamber door while the door is closed.
Replacement of the door seal ring
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:
Press in the top and bottom of
the door seal.
Press in the left and right sides
of the door seal.
Press the remaining sections of
the seal.
Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.
The drain valves
1. Press the included hose
on to the drain valve firmly.
2. Pull the drain valve
outward to drain the tank.
3. Push the drain valve back
after draining the tank.