![ICANCLAVE STE-12-D Instruction Manual Download Page 22](http://html1.mh-extra.com/html/icanclave/ste-12-d/ste-12-d_instruction-manual_609605022.webp)
22
The time required for sterilizer to be ready for routine use after the power is switched is less than 15
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
(STE-12-D)
Tem
per
at
ur
e
(°C
)
Pr
essur
e
(kPa)
Ho
ldi
ng
ti
m
e
(m
in
)
Tot
al
ti
m
e
(m
in)
Type
Ma
x
lo
ad
(
kg
)
Ma
x
lo
ad
p
er
tr
ay
(k
g)
SOLID
134
210
4
15-20
Unwrapped solid material
3.00
1.00
121
110
20
30-40
WRAPPED
134
210
4
30-45
Unwrapped solid material
3.00
1.00
121
110
20
45-50
Single-wrapped solid or
hollow
material
2.50
0.80
TEXTILE
134
210
8
30-45
Unwrapped porous material
0.75
0.25
Single-wrapped porous
material
0.50
0.15
121
110
30
55-60
Dual-wrapped porous material 0.30
0.10
Single-wrapped hollow
material
2.50
0.80
Dual-wrapped solid and
hollow
material
1.20
0.40
PRION
134
210
18
40-55
Unwrapped porous material
0.75
0.40
Single-wrapped porous
material
0.50
0.15
Dual-wrapped porous material 0.30
0.10
Single-wrapped hollow
material
2.50
0.80
Dual-wrapped solid and
hollow
material
1.20
0.40
LIQUID
(optional)
134
210
10
35-55
Liquid
0.80
0.25
121
110
30
40-50
Drying
(optional)
―
―
―
1-20
―
―
―
B&D test
134
210
3.5
15-25
―
―
―
Helix test
134
210
3.5
15-25
―
―
―
Vacuum test
―
―
―
18-21
―
―
―