Actual temperature: input the measured standard temperature (temperature stable). Sensor
temperature: temperature measured by the sensor.
Timing mode: count up timing and count down timing. Timing point: start of heating
and reaching set temperature.
Restore factory setting: restore parameter setting for delivery, wipe customer storage schemes.
4.2.3.4
Press Enter key on the main interface for help module:
Fig. 18
5.
Maintenance and Repair
5.1.
Maintenance:
5.1.1.
Check the power cord and power supply regularly. Replace the damaged and aging components
timely.
5.1.2
Check the air pipelines, valves and connectors regularly. If air leakage or loose connection occur, or
any damaged or aging components are found, replace them timely.
5.1.3
Clean the Scrubber upper cover regularly (once a week will be the best). The specific cleaning steps:
5.1.3.1
Feed about 100ml distilled water into the digestion tube.
5.1.3.2
Put the digestion tube on the rack. Set up the exhaust receiver. Open the condensed water.
5.1.3.3
Choose the linear heating mode. Set the temperature between 120
℃
-150
℃
.
(The temperature may vary with the altitude and atmospheric pressure. It should make sure that the
water does not boil intensely.)
5.1.3.4
It should distill about 30 minutes. This process may repeat several times according to the degree
of contamination.