![IBRAMED SONOPULSE III Instruction Manual Download Page 20](http://html1.mh-extra.com/html/ibramed/sonopulse-iii/sonopulse-iii_instruction-manual_3285117020.webp)
20
Manufacturer’s guidelines and declaration – Electromagnetic immunity
Immunity
test
Electromagnetic environment – guidelines
Test level
IEC 60601
Conformity
level
RF Conducted
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
RF Radiated
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Communication equipment of RF portable and mobile
should not be used near any part of
SONOPULSE III
including cable with separation distances smaller than the
recommended, calculated from the equation applicable to
the transmitter frequency
Recommended separation distance
d = 1.2
d = 1.2 80 MHz to 800 MHz
d = 2.4 800 MHz to 2.5 GHz
Where P is the maximum nominal output power in watts
(W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m). It
is recommended that the field intensity established by
the RF transmitter, as determined by an electromagnetic
inspection on the local, be smaller than the conformity
level in each frequency range . Interference may occur
around the equipment marked with this symbol:
SONOPULSE III
is destined to be used in the electromagnetic environment specified below. The user of the
equipment should ensure that it is used in such environment.
ELECTROMAGNETIC COMPATIBILITY GUIDANCE
P
P
P