7.1.4 TEMPERATURE COMPENSATION
Listed below are the temperature coefficients used by most major hemodialysis machine and reverse osmosis manufacturers.
We recommend that you double check this data with documentation from your machine’s manufacturer.
Display
To Use For
Temperature Coefficient
Temperature Coefficient
at
25°C
at
37°C
BAXTER
Baxter (Europe Dialysis Machine)
2,200 %/K
2,200 %/K
BELLCO
Bellco
2,000 %/K
2,100 %/K
B.BRAUN
B.Braun
2,000 %/K
2,100 %/K
FRESENI
Fresenius
2,000 %/K
2,100 %/K
GAMBRO
Gambro
2,000 %/K
2,070 %/K
AK-Type and Phoenix Machines
HOSPAL
Hospal
2,000 %/K
2,070 %/K
NIK ALL
Nikkiso Overall Conductivity
2,050 %/K
2,050 %/K
NIK BIC
Nikkiso Bicarbonate Conductivity
2,200 %/K
2,200 %/K
STANDARD
Other Dialysis Machines
2,000 %/K
2,070 %/K
USER TC
Adjustable
0 ... 5,000 %/K
0 ... 5,000 %/K
nLF-ISO
Non Linear Function for
According Standard
tap, purified and waste water.
ISO7888 – DIN EN27888
Reference
temperature
25°C
RO-DWA
Reverse Osmosis DWA
0,00 %/K
RO-DWA
DWA GmbH & Co. KG
Ubstadter Straße 28
76698 Ubstadt-Weiher
RO-HERCO
Reverse Osmosis Herco
2,200 %/K
RO-HERCO
Herco Wassertechnik GmbH
Reference temperature 25°C
Planckstraße 26
71691 Freiberg
RO-LAUER
Reverse Osmosis Lauer
nLF-ISO according Standard
RO-LAUER
LAUER MEMBRAN WASSERTECHNIK
ISO7888 – DIN EN27888
GmbH
Reference temperature 25°C
Industriegebiet Speichermatt
79599 Wittlingen
RO-DWA
Reverse Osmosis DWA
0,00 %/K
RO-DWA
DWA GmbH & Co. KG
Ubstadter Straße 28
76698 Ubstadt-Weiher
RO-HERCO
Reverse
Osmosis
Herco
2,200
%/K
RO-HERCO
Herco Wassertechnik GmbH
Reference temperature 25°C
Planckstraße 26
71691 Freiberg