12. TROUBLESHOOTING
This is a short troubleshooting guide. See service manual for full troubleshooting guide.
Problem
Verification
Cause
Corrective
No movement is
working.
Fast beep with green
light off on control box.
The battery is low and
equipment
is
not
connected to the wall
outlet.
Connect the equipment to
the wall for 8 hours. Once
plugged
in,
try
the
movements.
If
problem
persists, go to next cause.
(
CAUTION: Never allow
the battery to discharge
completely, as this may
damage the battery
).
No audible alarm and
green light off on the
control box.
Equipment not connected
to the wall outlet.
Connect equipment to wall
outlet.
If
the
problem
persists, continue to the
next cause.
Wall outlet is defective.
Make sure there is power in
the electrical outlet. If the
problem
persists,
the
control box is defective.