6.1 Chair functions
WARNING.
Always ensure there is sufficient clearance around chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING.
Always lock wheels before the patient is placed on chair.
WARNING.
Always lower chair to minimum height for patient access.
WARNING.
Always have patient enter and exit chair from the front.
WARNING.
Do not leave patient unattended.
WARNING.
Once arm and backrest are locked in upward position, make sure they
are securely locked.
WARNING.
Ensure that lateral movement is locked after use and before patient
sits in the chair.
WARNING.
Avoid placing hands and other body parts near mobile parts or locking
mechanism.
Lever allowing lateral sliding.
Pull lever towards back of the
chair, slide and release lever
to lock.
Foot pedal for height adjustment:
•
Raise: pump down lever.
•
Lower: push down lever at its lowest
position and hold.
•
Lock seat rotation: lift lever up.