OPERATION MANUAL-ECHO-FLEX
™ 5002
© Copyright IBIOM Instruments ltd 2021
Page 9 of 31
Revision: January 2021
6.2 Installation
WARNING.
Always ensure there is sufficient clearance around the table before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING.
ECHO-
FLEX™ table is designed for indoor use on flat surfaces.
WARNING.
Always lock wheels before the patient is placed on table.
WARNING.
Never use braking system to stop a moving table.
WARNING.
Do not leave patient unattended.
WARNING.
Do not use in oxygen-rich environment.
WARNING.
Do not use in MRI environment.
WARNING.
Do not use in wet environment.
WARNING.
Do not use for CPR or X-ray procedures or with defibrillator.
WARNING.
Do not roll on power cable, hand or foot control cable.
WARNING.
Keep all cables out of reach of children.
WARNING.
Keep all cables away from moving parts.
WARNING.
Do not move table by pulling hand control, power cable or any other
cable.
WARNING.
Never use the table if the power cable or any other cables are
damaged.
WARNING.
Keep the power cable out of heat sources.
WARNING.
Never exceed duty cycle of 2/18 (2 minutes ON and 18 minutes OFF).
ECHO-
FLEX™ table is delivered fully assembled and ready for use. The table must be unpacked
and inspected upon receipt to ensure that there has been no damage during transport.
1. Position table to desired location.
IMPORTANT
: Make sure you have enough space
around table to avoid collisions between its surroundings in ALL POSITIONS. Dimensions
show maximum space needed. Dimensions may be less than the one indicated below
depending on options.