ibidi Glass Bottom Dish Instructions Download Page 2

Instructions

Glass Bottom Dish

35 mm

Seeding Cells

Depending on your cell type, application of a 4–9 ×
10

4

cells/ml suspension should result in a confluent layer

within 2–3 days.

• Trypsinize and count cells as usual. Dilute the cell

suspension to the desired concentration.

• Apply 400 µl cell suspension into the inner well of

the µ–Dish. Avoid shaking as this will result in inho-
mogeneous distribution of the cells.

• After cell attachment add additionally 1.6 ml of pure

medium to ensure optimal grow conditions.

• Cover the µ–Dish with the supplied lid. Incubate at

37°C and 5 % CO

2

as usual.

We recommend not to fill more than the indicated to-
tal volume into the Glass Bottom Dish

35 mm

in order to

avoid the liquid contacting the lid.

Undemanding cells can be left in their seeding medium for
several days and grow to confluence there. However, best
results are achieved when the medium is changed every
2–3 days. Carefully aspirate the old medium and replace
it by up to 2 ml fresh medium.

Glass Bottom Dish Selection Guide

Glass Bottom
Dish

35 mm

µ–Dish

35mm, high

Glass Bottom

Lid

normal

with Lid Lock po-
sition

Evaporation

medium

low

Coverslip

No. 1.5

No. 1.5H

Thickness

170 µm
(+20 µm/-10 µm)

170 µm (±5 µm)

Packaging

10 pieces

1 piece

Note:

If cultivated in a dry atmosphere, evaporation can
harm the cells during longer experiments. Exchange
the medium every 1-2 days to minimize evaporation-
mediated osmolality shifts.

Optional Glass Coverslip Cleaning Protocol

The Glass Bottom Dish

35 mm

is made with an uncoated

glass coverslip. For improved cell attachment, glass sur-
face modifications and other applications, the glass cov-
erslip of the Glass Bottom Dish

35 mm

can be cleaned by

following the protocol below.

• Remove lids and immerse products in ddH

2

O in an

appropriately sized beaker.

• Sonicate for 10 minutes.

• Decant the ddH

2

O completely.

• Add 1 M HCl.

• Sonicate for 10 minutes.

• Decant the HCl completely and wash twice with

ddH

2

O. Decant the ddH

2

O completely.

• Add 2-propanol (absolute).

• Sonicate for 10 minutes.

• Aspirate the 2-propanol completely. Make sure that

all products are completely dry. Wash twice with
ddH

2

O and aspirate the ddH

2

O completely.

• Add ethanol (absolute).

• Sonicate for 10 minutes.

• Aspirate the ethanol completely.

Make sure that

all products are completely dry. Wash twice with
ddH

2

O.

• Sonicate in ddH

2

O for 10 min.

• Decant ddH

2

O and blow dry carefully with canned

air or clean nitrogen gas.

Modifications of this protocol including acids, bases, alco-
hols and detergents are possible. Please check the chem-
ical compatibility list on

www.ibidi.com

for compatibil-

ity. Make sure to handle the glass-bottomed products with
care. The glass coverslips may break during mechani-
cal handling. For best results, use a custom-made Teflon
holder.

Glass Bottom Dish