ibidi Heating System 4 Slides – Silver Line
7
Maintenance
7.1
Disinfection and Cleaning
The heated components and the ibidi Temperature Controller – Silver Line do not have to be sterile,
because the incubation chamber has no direct contact with the cells and the cell culture medium. If
disinfection is necessary for some reason, we recommend using isopropanol (70%) or common lab
disinfection solutions based on quaternary ammonium compounds (e.g., Barrycidal 36 or Pharmaci-
dal).
All parts of the ibidi Heating System 4 Slides can be cleaned from the outside. We recommend using
ultrapure water for cleaning. Fingerprints on the Heated Lid can be removed using isopropanol
(70%) or lens/eyeglass cleaning wipes.
CAUTION
– When cleaning the heated glass plate(s), be careful when wiping the
inner surface of the lid not to damage the electro-conductive coating. Also take care not
to damage the glass plate.
The use of ethanol or other types of organic solvents may remove the instrument’s paint.
To clean the heated components and/or the Temperature Controller switch off the ibidi Temperature
Controller – Silver Line and disconnect all electrical cables. Leave the instrument to cool down for
approx. 5 minutes.
WARNING
– Hot surface (max. 55°C)! Do not touch Heated Lid, Heated Plate,
and Heated Glass Bottom when hot.
7.2
Influence of Ambient Temperature and Ventilation
The ambient temperature affects the conditions inside the incubation chamber. Devices, such as
computers and camera controllers, can significantly heat up small rooms. In this case, we recommend
equilibrating the room temperature to the typical experimental conditions at least 2–3 hours before
starting the experiment.
Ventilation can enhance the effect of temperature and humidity changes in the vicinity of the incuba-
tion chamber. In a case where the airflow (e.g., air conditioning) cannot be stopped, we recommend
protecting the microscope as much as possible.
The use of an XL-Incubator and/or an objective heater minimizes those effects and helps significantly
to stabilize surrounding conditions.