Technical features
Characteristics
Specifications
Power supply
Input: 100-240V, 47-63Hz, 1.35A
Output: 15V DC, 3A max
Dimensions: 143x73x40mm
Classification
(EN 60601-1)
Class I
Applied part
(EN 60601-1)
Type BF
Dimensions
(length x height x
depth)
250x82x185mm
Output power (±20%)
0.5W-10.0W, when duty cycle
≥
80% for 5cm²
ultrasound head
0.5W-15.0W, when duty cycle
≤
70% for 5cm²
ultrasound head
0.1W-2.0W, when duty cycle
≥
80% for 1cm²
ultrasound head
0.1W-3.0W, when duty cycle
≤
70% for 1cm²
ultrasound head
Ultrasound wave
frequency
1MHz ±10%, 3MHz ±10%
Duty cycle
10%-100%, stepping 10%
Working frequency
100Hz ±10%
Therapy time
Adjustable, max 30 minutes
Timer accuracy
±3%
Effective radiating
area (Aer)
1.0cm² ±20% (optional)
5.0cm² ±20%
Effective intensity
3.0W/cm² ±20%
Accuracy
±20% for each setting above 10% of the maximum