18
MNPG117-02
2.
Ask about his/her sensation during treatment. If necessary, adjust
ultrasound intensity, by reducing it if the treatment is not
comfortable.
3.
In case of indications of wrong contact, it is recommended to add the
contact gel or reposition the ultrasound-head.
4.
During the treatment if the ultrasound head works correctly, the
applicator LED will light; if there is no contact, the applicator LED will
blink light. When the treatment is in PAUSE, the applicator LED will be
turned off and the countdown will also be stopped.
WARNING:
→
The treatment should be performer with a regular movement of
the ultrasound head, not too slow to avoid inducing heat, not too
fast to prevent a bad contact which would reduce the
effectiveness of the treatment.
→
If it is needed to replace the handle, turn the power switch off
and disconnect the device from power supply.
After the treatment clean the skin of the treated area as well as the ultrasound
head by using a dry towel. The ultrasound-head should be cleaned up with a
70% alcohol solution. Check the patient conditions and the treated area (pain,
circulation, etc.).
The patient should reveal any complaint/reaction before starting the
treatment after.
Ultrasound treatment
After performing the preliminary operations listed in the previous paragraph,
start the session making sure to follow these steps:
1.
Press
PROGRAM
keys
to select the program: scroll up/down
the programs with the arrows.
2.
Select the frequency 1 or 3 MHz by pressing the button
FREQUEN.
.
3.
Select the ultrasound head
1 CM² or 5 CM²
by pressing the button
.
ATTENTION:
The user can select the ultrasound head 1 cm
or 5 cm only if both heads are connected. The device recognizes