IACER Srl
42
MNPG151-08
Guidance and manufacturer’s declaration – ELECTROMAGNETIC EMISSIONS –
FOR ALL EQUIPMENT AND SYSTEMS
MIO-IONOTENS is intended for use in the electromagnetic environment specified
below. The customer or the user of MIO-IONOTENS should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
MIO-IONOTENS uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
MIO-IONOTENS is suitable for
domestic establishment and in
establishment directly connected to
the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonics emissions
IEC 61000-3-2
Class A
Voltage
fluctuation/flicker
emissions
IEC 61000-3-3
Compliant