INTENDED USE  

I-TECH  LA500  is  an  electro-medical  device  that  delivers  treatments  of  laser-therapy, 
with the help of  power laser up to 500mW for the provision of treatment though a 
specific probe. 
  
I-TECH  LA500  is  an  active  therapeutic  device,  not  invasive,  used  especially  by 
physiotherapists, physicians and pain therapists.  

The use of I-TECH LA500 is indicated for professional user in clinics/hospitals and for 
home users in home environment.   

The operator, in fact, should be qualified to be able to use such equipment, and he 
should  have  passed  an  adapted  training,  or  should  operate  under  the  control  of  a 
medical  adequately  qualified  to  the  use  of  the  equipment,  in  order  to  guarantee 
safety conditions to the patient. 

Such  equipment  can  be  used  in  hospital  environment  outpatient,  nevertheless,  it  is 
important  to  know  that  the  user  follows  the  medical  instructions  to  use  the 
equipment or that he follows the indications present in the user’s manual. 

INDICATIONS  

The  application  fields  that  can  profit  from  using  the  laser-therapy  I-TECH  LA500 are 
the following: 

 

1. Arthro-rheumatic pathology  
Arthrosis, sciatica, scapular and humeral periarthritis, arthropathy of hands and feet, 
epicondylitis,  hip  arthrosis  at  its  initial  stage,  gonalgia  with  or  without  effusion, 
myogenic stiff neck, lumbago, myositis, chronic and acute pathologies etc.  
 
2. Rehabilitative therapy  

Articular motor rehabilitation after removing plaster apparatuses or after orthopaedic 
surgical operations.  

3. General medicine and dermatology  
Decubitus  ulcers,  cheloids,  torpid  sores  for  its  well-known  bio-stimulating  and  anti-
infectious effects.  
 
 

CONTRA-INDICATIONS  

-

 

direct radiation in the eye: the human eye is extremely sensitive to laser radiation 
and  can  be  permanently  damaged  from  direct  or  reflected  laser  beams.  The 
special safety glasses must be worn by both patient by the operator.  

-

 

pregnancy:  the  laser  is  contraindicated  for  use  over  the  pregnant  uterus.  It  can 
still  be  used  in  pregnant  women  with  the  foresight  to  not  radiate  over  the 
abdomen. 

-

 

neoplasms:  you  must  not  use  the  laser  on  a  primary  lesion  or  secondary  non-
diagnosed.  Laser  treatment  may  be  granted  to  relieve  pain  during  the  terminal 
stage  of  disease,  it  is  recommended  that  this  be  performed  only  with  the  full 
consent of the patient. 

-

 

thyroid: the laser should not be used in any case above the gland. 

-

 

bleeding: vasodilatation may worsen the bleeding. 

-

 

immuno-suppressive  therapy:  laser  therapy  is  contraindicated  in  patients  who 
have undergone such type of drug therapy. 

-

 

in the skin and injuries suspects: never laser radiation on angiomas, black points 
or injuries suspects on the skin. 

-

 

 treatments  over  the  sympathetic  ganglia,  the  vagus  nerve  and  region  of  the 
heart  in  patients  with  heart  disease:  laser  therapy  can  significantly  alter  the 
function  neural,  and  is  therefore  contraindicated  in  this  regions  body  patients 
with heart disease. 

Other: 

-

 

Atopic dermatitis and eczema in the acute phase 

-

 

Inflammatory processes in place at the site to be treated 

-

 

Abrasions or bruises 

-

 

Photoallergy 

-

 

Recent surgery or cryotherapy in cutaneous sites to treat 

WARNING  

-

 

Photosensitivity  reactions:  in  some  patients  taking  drugs  known,  the  latter 
are  due  to  reactions  of  photosensitization.  It  is  not  clear  how  the 
combination  of  laser  and  drugs  triggers  this  response.  It  is  recommended 
that patients at risk allergic, or patients with a history of such reactions  are 
"tested" with a minimum time of treatment. 

-

 

Means  of  fixation,  metal  plates,  plastic  NOT  constitute  contraindication  to 
the use of lasers which can be safely used on metallic implants, sutures and 
plastics. 

Summary of Contents for LA500

Page 1: ...I A C E R Srl MNPG127 02 06 06 17 LASER THERAPY...

Page 2: ......

Page 3: ...E 8 INDICATIONS 8 CONTRA INDICATIONS 8 DEVICE DESCRIPTION 9 PATIENTS PREPARATION 9 HOW TO USE IT 9 Therapeutic indications 11 MAINTENANCE 13 HANDLES CLEANING 13 WORKING PROBLEM 13 ELECTROMAGNETIC INTE...

Page 4: ...important information and useful tips for operating the equipment CAUTIONS The CAUTION message appears before operations which if not correctly performed may cause damage to the machine and or its acc...

Page 5: ...t use the equipment in places where it might get wet Before operating the machine carefully check the correctness of the connections according to the instructions Do not use accessories other than the...

Page 6: ...ecution of different procedures from those specified in this user manual can cause the exposition to dangerous radiation The operator has the responsibility to verify that the issuing head remains wel...

Page 7: ...er production of ATP The applications of the I TECH LA500 laser produce a number of effects on the treated tissues 1 increase in hematic flow vasodilatation of capillaries and arteries 2 biostimulatio...

Page 8: ...ensitive to laser radiation and can be permanently damaged from direct or reflected laser beams The special safety glasses must be worn by both patient by the operator pregnancy the laser is contraind...

Page 9: ...of the patient for the therapy carry out the following operations 1 wash the skin using soap and mild soap or alcohol where the head of the laser will be positioned 2 dry well the skin HOW TO USE IT...

Page 10: ...ent time last 3 minutes Select the operation mode CONTINUOS or CYCLE using the buttons on the right side of the front panel picture below With CYCLE mode it is possible to select the ON time from 1 to...

Page 11: ...ition upper picture Take away the key from the machine to avoid the use by unauthorized personnel Please do not switch off the device unplugging the cable from the power socket to avoid any type of da...

Page 12: ...ITIS 0 3 66 sec 264 sec BAKER S CYST 0 3 66 sec 264 sec DE QUERVAIN SYNDROME 0 2 100 sec 400 sec OEDEMA 0 3 66 sec 264 sec HEMATOMA 0 3 66 sec 264 sec EPICONDYLITIS 0 4 50 sec 200 sec HERNIATED DISC 0...

Page 13: ...ent HANDLES CLEANING The laser handle is very weak and it needs a daily cleaning The following advices are absolutely important to not damage the lens and the laser probe It is suggested to 1 Remove t...

Page 14: ...that the socket works properly The power cable is not inserted correctly into the connector of the device Insert correctly the plug and the cable in the connector of the device The power cable is wor...

Page 15: ...B Output channels 1 Class of isolation parts applied according to the rule EN 60601 1 I BF Protection level against the liquids penetrations according to the rule EN 60601 1 IPX0 Command of execution...

Page 16: ...that the product shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognized centre of recycling of electrical and electronic equipment Appendix B...

Page 17: ...ctrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Page 18: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Page 19: ......

Page 20: ......

Page 21: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...

Reviews: