0123 

 

 

 

 

 

 

 

 

 

 

Figure 3.5.1                                            Figure 3.5.2

 

The Items in Page1 are: 

Items 

Default 

Options 

Description 

Alm 

on 

on, off 

Turn on or off the sounds when the value 

if out of the limit set in page 2. 

Beep 

off 

on, off 

Set Pulse Beep on or off. 

Demo 

off 

on, off 

Enter Demo mode, long press to quit it. 

Reset 

OK 

OK, Blank 

Reset all settings as default value. 

Brightness 

1-5 

Set the screen

’s brightness. 

Exit 

N/A 

N/A 

Return  to  the  Main  Screen  or  Enter 

Demo mode when it is selected as 

‘on’. 

The Items in Page2 are: 

Items 

Default 

Options 

Description 

SpO2 Alm Hi 

100 

50-100 

The upper limit of SpO2. 

SpO2 Alm Lo 

94 

50-100 

The lower limit of SpO2. 

PR Alm Hi 

130 

25-250 

The upper limit of PR. 

PR Alm Lo 

50 

25-250 

The lower limit of PR. 

+/- 

+, - 

Increase or decrease the number. 

Exit 

N/A 

N/A 

Return to the Main Screen. 

3.6 Turn off the Oximeter 

The oximeter will turn off automatically after 8 seconds after removing the 

finger out of the device. 

Section 4    Maintenance

 

4.1    Cleaning 

Switch off the power and take out the batteries before cleaning, Cleaning 

exterior  surface  (OLED  display  screen  included)  of  the  unit  with  a  dry  and 

soft cloth. Use 75% density of medical alcohol to clean the surface and use 
dry fabric with little alcohol to avoid alcohol permeates into the device. 

4.2    Disinfection 

Use  75%  density  of  medical  alcohol  to  clean  the  surface  that  contacting 

with the user. 

CAUTION:

 

Don’t use strong solvent. For example, acetone.

 

CAUTION:

 Never use an abrasive such as steel wool or metal polish.

 

CAUTION:

 Do not allow any liquid into the product, and do not immerse any 

parts of the device into any liquids.

 

CAUTION:

 Avoid pouring liquids on the device while cleaning.

 

CAUTION:

 

Don’t remain any cleaning solution on the surface of the device.

 

 
4.3    Warranty 

The  host  product

’  design  life  is  2  years,  and  1  years  warranty.  Under 

normal  circumstances,  the  malfunction  of  the  product  during  the  warranty 

period (from the date of purchase) should  be sent back to the company for 
maintenance,  and  our  company  is  responsible  for  all  maintenance  costs 

(users should cover the freight themselves). Outside the warranty period, our 

company shall charge some maintenance fee (users should cover the freight 
themselves)  if  the  product  has  broken  down  and  is  sent  back  for 

maintenance. The battery is beyond the scope of the warranty. If you have 

the  purchase  and  sale  contract,  the  costs  of  the  maintenance  shall  be  in 
accordance  with  the  purchase  and  sale  contract  execution.  Besides,  it  is 

recommended that users should use them no more than five years. And over 

the using life, the using risks may increase due to the equipment

’s aging.

 

4.4    Maintenance 

 

Replace  the  batteries  timely  when  battery  indication  is  low.  Clean 

surface of th

е Pulse Oximeter before it is used in diagnosis for users. 

 

Remove the batteries inside the battery cassette if the Oximeter will not 
be operated for a long time. 

 

It  is  better  to  preserve  the  product  in  a  place  where  ambient 

temperature is -20 - 55

 and humidity is 10%-95%. 

 

Regular  inspection  to  make  sure  that  no  obvious  damage  existed  to 

affect the safety and performance of device. 

 

No  flammable  substance,  overtop  or  lower  temperature  and  humidity 
existed in operation conditions. 

4.5    Troubleshooting

   

Table 4.5    troubleshooting 

Problems 

Possible Reason 

Resolutions 

Oxyhemoglobin 
or heart rate 
can not be 
shown 
normally. 

1. Finger is not plugged correctly. 

2. User

’s perfusion is too low to 

be measured. 

1. Retry by plugging the finger. 

2. Try some more times, if you can 
make sure about no problem 
existing in the product, Please go 
to a hospital timely for exact 
diagnosis. 

Oxyhemoglobin 
or heart rate is 
shown 
unstably. 

1. Finger might not be plugged 
deep enough. 

2. Finger is trembling or user's 
body is in movement status. 

1. Retry by plugging the finger. 

2. Try not to move, Let the user 
keep calm. 

The oximeter 
can’t go into the 
working state. 

1. Power of batteries might be 
inadequate or not be there at all.     

2. Batteries might be installed 
incorrectly.   

3. The Oximeter might be 
damaged. 

1. Please replace batteries.   

2. Please reinstall the batteries. 

3. Please contact with local 
customer service center. 

The screen are 
suddenly off. 

1. The product is automatically 
standby or sleep when no signal 
is detected longer than 8 
seconds. 

2. Power quantity of the batteries 
is exhausted. 

1. Normal. 

2. Replace the batteries. 

4.6    Disposal   

To  avoid  contaminating  or  infecting  personnel,  the  environment  or  other 

equipment,  make  sure  you  disinfect  or  decontaminate  the  device 
appropriately before disposin

g of it in accordance with your country’s law for 

equipment containing electrical and electronic parts. 

 

Section 5    Specification

 

Physical Characteristics 

  Machine Dimensions: 57mm (L) x 31mm (W) x 30.5mm (D) 
  Machine Weight -approx: 54 g ( including 2xAAA battery ) 

Classification

 

  Anti-electric Shock Type: Internally powered equipment 
  Anti-electric Shock Degree: Type BF equipment 

  EMC: Group 1 Class B 

  Mode of operation: Continuous Operation 
  Enclosure Degree of ingress protection: IP22 

 

IP22 means shell  of  this  product  can withstand  the water  dropping to 

the surface when the shell deviate 15 degree from horizontal surface.

 

Power 

Internal: 

2xAAA 1.5V alkaline battery 

Power Consumption: 

30mA(Normal) 

Environmental: 

Operating Temperature: 

5°C to 40°C 

Storage Temperature: 

–20°C to 55°C 

Relative Humidity: 

15% to 85% non-condensing 

Electronics Parameters: 

Parameter

 

Value

 

SpO2(Oxygen saturation) 

35-100%   

PR (Pulse Rate) 

25-250 bpm 

Resolution 

SpO2 (Oxygen saturation) 

1% 

PR (Pulse Rate) 

1 bpm 

PI (Perfusion Index) 

0.1% (<10%) 
1% (10% - 20%) 

Measure 
Accuracy: 

A

rms

 * 

SpO2(Oxygen saturation) 

2% (80% - 100%) 
3% (70% - 80%) 
Unspecified (<70%) 

PR (Pulse Rate) 

2 bpm 

PI (Perfusion Index) 

1% 

A

rms

  accuracy  is  a  statistical  calculation  of  the  difference  between  device 

measurements  and  reference  measurements.  Approximately  two-thirds  of 
the device measurements fell /- Arms of the reference measurements 

in a controlled study.

 

  Applicable models 

Technical information is subject to change without notification to allow for updates. 

  This device complies with EU Directive 93/42/EEC concerning medical devices, and 

the DIN EN ISO 13485:2016 standard. 

  This device complies with IEC 60601-1-2: 2014, ISO 80601-2-61: 2011 Clause 202, 

IEC 60601-1-11: 2010 Clause 12.   

  Manufacturer: Promise Technology Co., Ltd. 

3/F, East-Asia Building, Jida Jiuzhou Avenue, 519015 Zhuhai, Guangdong, 

PEOPLE

’S REPUBLIC OF CHINA 

  EC-Representative: Shanghai International Holding Corp. Gmbh (Europe)   

Eiffestraβe 80 20537 Hamburg GERMANY 

Distribuited by: I.A.C.E.R Srl Via S.Pertini 24/a 

– 30030 Martellago (VE) Italy 

Settings 

Page 1/2              * 

Alm  

 

            on 

Beep 

 

            off 

Demo 

 

 

    off 

Reset 

 

 

    OK 

Brightness 

 

      4 

Exit 

Settings 

Page 2/2                  * 

SpO2 Alm Hi                100 

SpO2 Alm Lo                  94 

PR Alm Hi 

 

      130 

PR Alm Lo 

 

        50 

+/-                       

 

  + 

Exit 