Error messages
34
Error
description
Cause
Remedy
If this error occurs repeatedly, send the blood pressure monitor to
your healthcare provider or directly to the manufacturer for
checking.
Err 5
Power pack or battery
voltage too low.
Change the batteries (
Chapter 5.2
).
Battery contacts are
corroded.
Clean the battery contacts with a cotton cloth and a little alcohol.
Err 6
Air congestion
Check the blood pressure cuff for air congestion or a kink in the
air hose. If there is a kink in the air hose, unkink it, otherwise send
the blood pressure monitor to your healthcare provider or the
manufacturer.
Blood pressure cuff
incorrectly connected.
Connect the blood pressure cuff to the blood pressure monitor
(
Chapter 6.2
).
Leak in the blood
pressure cuff or air hose
Replace the blood pressure cuff with the air hose.
Abr. (Abort)
Measurement cancelled
on request
Do not press this button when measuring, unless you want to
abort the measuring.
Err 9
Residual pressure in the
blood pressure cuff
Wait until the blood pressure cuff has deflated completely.