Configuration
§
Intelligent
: Logger detects when a receiver is in range and then sends data automatically in the measuring
interval set.
§
Always
: The logger will always send data, not matter whether there is a receiver in range or not. This setting
requires continuous power!
§
Off
: The logger does not send any data.
§
HK-Counter
: Set transmission interval of internal self-monitoring values. 1 = with every normal transmission. 10 =
with every 10th transmission. In case of alarm the self-monitoring values will be collected at each measurement.
§
Log HK
: Select, which self-monitoring values shall be collected. The following options can be selected: Supply
(battery), temperature (Housing temperature) and humidity (Housing inside humidity).
§
Parameter Set
: Shows which set of parameters is currently selected and cannot be changed. The number contains
the logger ID, year, month, day, hour, minute, second (e.g. 100012_20090812080050: Logger Nr 100012 _ 12.
August 2009 at 08:00:50) The number is set every time parameters are sent to the logger.
§
GPRS / GSM Setup
: Opens the window for transmission and modem settings.
§
Channel Parameters: Settings of the channel parameters can be changed here. Channel 1 is for the measurement of
water level / tapping, channel 2 is for measuring the water temperature.
§
# 1
: means channel 1 is selected. Switch the channel to be edited by clicking the arrow up / down buttons.
§
Type
: indicates which value is measured in this channel and cannot be changed manually.
§
Unit
: Choose the Unit to be measured.
§
ID
: This parameter is intended to mark the measuring point, for example, in a database. As long as "0" is set, this
value has no effect. Values from 1 to 65535 can be entered. The value entered here will be listed in the table
headings of the exported data. Whether and how this information will be taken into consideration when importing
the data is the responsibility of the importing software.
§
Action
: Each action can be activated by checking.
§
Log channel
: Measured data of this channel will be saved when this option is checked.
§
Check alarms
: Check this option if you want to set special alarm events (e.g. shorter measuring and
transmissions intervals).
3.1.5
Scale
§
Offset
: Offset (available if no pressure sensor is used for the water level): This value will be subtracted from the
measured value (e.g. if the value should be displayed as height above sea level).
Note
: positive numbers are
subtracted from the measured value, negative numbers are added!
§
Multi
: Default setting is 1.000000. If a minus (-) is set before the first digit, the value entered becomes negative. Use
this to select whether water column or tapping should be measured. The adjustment value can also be calculated
under the tare button. Furthermore, the density of the water can be set here. The value 1.000000 is calibrated for
fresh water. Depending on the salinity, the value has to be reduced (i.e.: 3% salt content = 0.997000).
§
Tare
).
§
Alarm
: If the thresholds entered here are exceeded or fallen below, the device will switch to alarm mode. High: If the
measured value falls below the value entered here, the alarm is activated.
§
Low
: If the value entered here is reached or exceeded, the alarm is activated.
§
Power Up Wait
: warm-up time for the sensor (enter 1000 ms for PTM/N from STS)
§
Sensor min / max.
: Not required if a pressure sensor is used. If slower sensors are used, a period of time can be set for
the measuring time. Settings depend on the sensor type.
§
Type
: Drop Down Menu. Choose between SDI (for general SDI sensors i.e. temperature), SDI12-Pressure (for the
pressure channel of the iLevel-GW and wireless sensor (to connect more sensors via near field radio)
§
No measure, use cached values
: must be enabled if a sensor sends multiple values (i.e. pressure and temperature)
§
Index
: Numbers of sensor channels (i.e. pressure = 0 as first channel and temperature = 1 as second channel.
§
S.Nr.
: SDI 12 address of the selected sensor (default is 0)
§
Cmd.
: 0 means SDI command m!, which triggers measuring
3.1.6
More Buttons
§
Transfer
: Transfers the new Parameter Set to the logger.
§
Cancel
: Cancel without transferring the new parameter set to the logger.
§
Write Par...
: Save the new parameter set to the local computer without transferring it to the logger.
§
Read Par...
: Read a parameter set that has been saved to the local computer before.