Recommended Separation Distances

Recommended separation distances between portable and mobile RF 

communications equipment and the Hyperice X electronics box.

Rated maximum 

output power of 

transmitter W

Separation distance according to frequency of transmitters in 

meters

150 kHz – 80 MHz 

d=1.2/V1]√P

80 MHz to 

800MHz 

d=1.2/V1]√P

800 MHz to 2.5 GHz 

d=2.3√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the 

recommended separation distance (d) in meters (m) can be estimated using 

the equation applicable to the frequency of the transmitter, where P is the 

maximum output power rating of the transmitter in watts (W) according to 

the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher 

frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, 

objects, and people.

Equipment Classification

• 

Protection against electric shock: Class II/internally powered equipment

• 

Degree of protection against electric shock: Type BF applied part 

(Hyperice X device)

• 

Ingress protection: IP22

• 

Equipment not suitable for use in the presence of a flammable anesthetic 

mixture with air or with oxygen or nitrous oxide

• 

Continuous operation

Troubleshooting

Problem

Possible Causes

Solution

I cannot start 

the device

Power is not turned on.

Device battery is dead.

Not connected to app.

Press the power button to turn 

device on.
Plug the device into the charger 

to charge the battery.
Download the Hyperice App 

and pair your Hyperice X device.

The straps are 

too short.

Extenders are not installed.

Open the “Extender Box” 

and follow the instructions in 

the box.

The straps are 

too long.

Minimizers are not installed.

Open the “Minimizer Box” and 

follow the instructions.

Device does 

not get cold 

enough.

Adjust temperature setting 

within app
Cold therapy expectations are 

too extreme.
Air flow to modules is being 

blocked.
The device has been damaged

The device is sustaining 

optimum therapy temperatures, 

which is NOT ice cold.
Do not use device under 

pants, blankets, or anything 

that may restrict air flow to the 

device. If modules appear full 

of debris, attempt to clear with 

compressed air WHILE DEVICE 

IS NOT IN USE.
Contact Customer Service. Do 

not attempt to disassemble or 

fix device.

Summary of Contents for Hyperice X

Page 1: ...Operating Instructions HypericeX ...

Page 2: ...ay cause the device to operate incorrectly To avoid risk of strangulation do not leave a baby or child unattended with the power supply Choking hazard small parts Keep away from small children Do not leave the device power supply or any accessories where they could be damaged by children pets pests or liquids If you suspect your electronics box is damaged contact customer service at 1 949 565 4994...

Page 3: ...o not bleach On product tag Do not iron On product tag EU RF transmitter symbol In manual The Bluetooth figure mark On product Warning symbol to identify a hazard that may lead to death or serious injury In manual and on product Caution symbol to indicate the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for ...

Page 4: ...tronics box on your thigh Align colored ring to kneecap Step 2 Fasten straps so the device is snug against the skin See Troubleshooting if straps are not sized correctly Step 3 Download the free Hyperice App on an iOS or Android device Open the Hyperice App create an account and pair your device Step 4 Press the power button on the device Status light will pulse green Connecting to the Hyperice Ap...

Page 5: ...formation regarding available replacement parts and accessories Technical Information Do not attempt to take apart the device The device has no user serviceable parts There are no user replaceable fuses Bluetooth Wireless Technology The Bluetooth word mark and logos are owned by Bluetooth SIG Inc and any use of such marks by Hyperice is under license In the unlikely event of loss of a stable Bluet...

Page 6: ...ovide power for approximately 90 minutes of continuous use of cold therapy It takes approximately three hours to fully charge the battery when the device is plugged in and not in use The rechargeable lithium ion battery is intended to be changed only by authorized service personnel with the use of a special service tool Product Specifications Hyperice X dimensions 12 5 x 10 x 1 5 Hyperice X weight...

Page 7: ...magnetic environment specified below The user of this equipment should assure that it is used in such an environment Immunity Against IEC 60601 1 2 Test Level Compliance Level of this electronics box Electromagnetic Environment Electrostatic discharge ESD IEC 61000 4 2 Contact Discharge 8 kV Air Discharge 15 kV 8 kV 15 kV Floors should be wood concrete or ceramic tile If floors are covered with sy...

Page 8: ...of the equipment including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter as below Recommended separation distance d 1 2 V1 P d 1 2 P 80 MHz to 800MHz d 2 3 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separ...

Page 9: ... equipment Degree of protection against electric shock Type BF applied part Hyperice X device Ingress protection IP22 Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Continuous operation Troubleshooting Problem Possible Causes Solution I cannot start the device Power is not turned on Device battery is dead Not connected to a...

Page 10: ...POSE OR USE AND ARE LIMITED IN DURATION AS STATED ABOVE EXCEPT AS EXPRESSLY STATED ABOVE HYPERICE SHALL HAVE NO LIABILITY OR RESPONSIBILITY TO ITS CUSTOMER OR ANY OTHER PERSON OR ENTITY WITH RESPECT TO ANY LIABILITY LOSS OR DAMAGE CAUSED DIRECTLY OR INDIRECTLY BY USE OR PERFORMANCE OF THE PRODUCT OR ARISING OUT OF THE USE OR INABILITY TO USE THE PRODUCT OR ANY BREACH OF THESE WARRANTIES INCLUDING ...

Page 11: ... notified Unacceptable returns and or exchanges may be reshipped to you following payment of an inspection shipping fee Returned items will be subject to a 7 5 restocking fee A returned item is an item that is being sent back for a refund If you refuse delivery of your order for any reason you will be refunded the cost of your order less shipping fees All partial or full refunds will be posted to ...

Page 12: ...ce Inc All rights reserved 60 00525 Rev 2 This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation ...