INS0039  EN 

SEP 2019 

Rev. D 

UltraSnap™ Surface ATP Test 

For use with Hygiena™ ATP Monitoring Systems 

Part No: US2020 (100 tests)

Description/ Intended Use: 

UltraSnap Surface ATP Test is a self-contained device for 

use with Hygiena luminometers. The test device and 

luminometer create a system used for monitoring hygienic 

status of surfaces on processing equipment and other 

environments in a wide range of industries. The system 

works by measuring adenosine triphosphate (ATP), the 

universal energy molecule found in all animal, plant, 

bacterial, yeast, and mold cells. Product residues from 

organic matter left on surfaces contain ATP. Microbial 

contamination on a surface contains ATP but typically in 

smaller amounts. After proper cleaning, all sources of ATP 

should be significantly reduced. When a sample is collected 

and ATP is brought into contact with the unique liquid 

stable Luciferase/ Luciferin reagent in the UltraSnap test 

device, light is emitted in direct proportion to the amount of 

ATP present in the sample. The luminometer measures 

generated light and reports results in Relative Light Units 

(RLU). RLU result provides information on the level of 

contamination within seconds. The higher the RLU number, 

the more ATP present, and the dirtier the surface.  It is 

important to note that UltraSnap is designed to detect 

invisible/trace amounts of residue. Overloading the swab 

with physical matter by swabbing a visibly dirty surface will 

inhibit the bioluminescent reaction and produce inaccurate 

results.  

For water samples such as Clean In Place (CIP) rinse water 

testing, use AquaSnap™ Water ATP Test Devices. Visit 

www.hygiena.com

 for more information.

 

Directions:  

Instructional Video:

 

www.youtube.com/HygienaTV

 

Before beginning testing, turn on luminometer. If 

luminometer has been programmed with test locations, 

select appropriate location before running test. 

1.

Allow UltraSnap to equilibrate to room temperature

(21 – 25 °C) before use. Holding swab tube firmly, 

twist and pull top of swab out of swab tube.

Condensation may be visible on inside of swab tube; 

this is normal.

2.

Thoroughly swab a standard 10 x 10 cm (4 x 4 inches)

area for a typical flat surface. Swab tip is pre-

moistened for maximum sample collection. For

irregular surfaces, ensure swabbing technique 

remains consistent for each test and swab a large

enough area to collect a representative sample.

Important swabbing technique tips:

Do not touch swab or inside of sample device 

with fingers.

Rotate swab while collecting sample to maximize

sample collection on swab tip.

Apply sufficient pressure to create flex in swab

shaft.

Swab in a crisscross pattern vertically,

horizontally, and in both diagonal directions.

Refer to instructional video for demonstration:

www.youtube.com/HygienaTV

 

3.

After swabbing, replace swab back in swab tube.

4.

To activate device, hold swab tube firmly and use

thumb and forefinger to break Snap-Valve by bending 

bulb forward and backward. Squeeze bulb twice,

expelling all liquid down swab shaft.

5.

Bathe swab bud in liquid by shaking

 

for 5 – 10 

seconds. Once activated, sample must be read in 

luminometer within 30 seconds.

6.

Holding luminometer upright, insert entire UltraSnap 

device into Hygiena luminometer. 

7.

A.

 If using the EnSURE™ Touch, close lid and press

“Run Test” to initiate measurement. Refer to the 

instrument manual for operating instructions. Results

will be displayed in 10 seconds.

B.

If using the EnSURE™ or SystemSURE Plus,

close lid and press "OK" to initiate the measurement. 

Refer to the instrument manual for operating 

instructions. Results will be displayed in 15 seconds.

Interpretation of Results: 

Hygiena luminometers are preset with Pass & Fail RLU 

limits of of 20 and 60 RLU for the EnSURE Touch and 10 

and 30 for the EnSURE and SystemSURE Plus. These 

limits are based on industry standards and published 

study recommendations. When using default settings, 

readings less than 20 on the EnSURE Touch (10 on 

EnSURE and SystemSURE Plus) RLU indicate surface is 

considered clean. Readings between 21-59 on the 

EnSURE Touch (11-29 on EnSURE and SystemSURE 

Plus) RLU indicate a warning, surface is not adequately 

clean. If reading is greater than 60 on the EnSURE Touch 

(30 on EnSURE and SystemSURE Plus) RLU, surface is 

considered dirty. Hygiena recommends setting RLU 

thresholds according to standards of your facility. Visit

 

www.hygiena.com

 

to view recommended practices or call 

our technical representatives for guidance. 

Calibration Control: 

It is advisable to run positive and negative controls 

according to Good Laboratory Practices. Hygiena offers the 

following controls:

 

(Part # PCD4000) Calibration Control Kit for Hygiena

Luminometers

(Part # CK25) ATP Positive Control Kit for ATP Test 

Devices

(Part # CAL) CalCheck LED Calibration Verification 

Device 

Storage & Shelf Life:

 

Store at 2 – 8 °C (36 – 46 °F) 

Test devices will tolerate temperature abuse for 4

weeks at room temperature (< 25 °C)

Store UltraSnap devices out of direct sunlight.

Devices have a 15-month shelf life. Refer to

expiration date on label.

Disposal: 

UltraSnap devices are made of 100% recyclable plastic 

and may be discarded accordingly.  

Safety & Precautions: 

Components of UltraSnap do not pose any health risk 

when used in accordance with standard laboratory 

practice and procedures of this insert. 

UltraSnap test devices are for one-time use. Do not

reuse.

For further safety instruction, refer to Safety Data Sheet 

(SDS). 

Hygiena Liability:

 

Hygiena will not be liable to user or others for any loss or 

damage whether direct or indirect, incidental or 

consequential from use of this device. If this product is 

proven to be defective, Hygiena’s sole obligation will be to 

replace product or at its discretion, refund the purchase 

price. Promptly notify Hygiena within 5 days of discovery 

of any suspected defect and return product to Hygiena. 

Please contact Customer Service for a Returned Goods 

authorization number. 