Hygenic Parabath Owner'S Manual Download Page 8

P A R A B A T H

®

7

IF YOU NOTICE

Paraffin hardening in 

the tank.

Paraffin has discolored.

Sediment in the tank.

Less than 3 inches of 

paraffin in the tank.

Operating light is out.

Paraffin takes longer than 

normal to melt.

THE LIKELY CAUSE

Frequent use has depleted 

the oil component of the 

paraffin.

Paraffin in use has a  

higher melting  

 

temperature than

Parabath

 

Paraffin.

Thermostat is not    

functioning correctly.

Normal use over time.

Normal use over time.

 

Normal depletion of 

paraffin, particularly in 

multi-person use.

Plug is loose or damaged.

Light is burned out.

Wax was frozen during 

shipment.

Unit was plugged in  

before it had completely 

cooled, preventing Quick 

Melt cycle from activating.

WHAT TO DO

Replace the paraffin. 

Replace the paraffin, using 

Parabath

 

Paraffin Refills.

Contact the Parabath

 

 

Product Dealer the unit 

was purchased from or 

The Hygenic Corporation 

to arrange for repair. 

Replace the paraffin.

Replace the paraffin.

Add additional paraffin 

sufficient to raise the level 

of paraffin to recommended 

levels.

Reconnect the plug. If  

damaged, seek qualified 

repair assistance.

Contact the Parabath 

Product Dealer the unit 

was purchased from or 

The Hygenic Corporation  

to arrange for repair.

Allow the unit to operate 

overnight before using.

Allow the unit to operate 

overnight before using.

T

roubleshooTing

 g

uide

Summary of Contents for Parabath

Page 1: ...Parabath Paraffin Heat Therapy Owner s Guide ATTENTION READ ALL INSTRUCTIONS BEFORE USING Keep manual in safe location for future reference...

Page 2: ...RATING UNIT Table of Contents Important Safety Information to be Read Before Using 2 Warnings 3 Instructions for Use 4 Treatment Steps for the ParabathUnit 5 Cleaning and Refilling 6 Troubleshooting G...

Page 3: ...ONS Do not use the ParabathUnit 1 When open cuts or broken skin are present 2 In the presence of inflamed or irritated skin conditions 3 When peripheral vascular disease or impaired circulation are pr...

Page 4: ...d use as described in the Parabath Owner s Manual The use of attachments not recommended by the manufacturer may cause hazards Use unit only with a verified 3 prong outlet that has been grounded to re...

Page 5: ...ontact with the bottom of the tank Plug in the electrical cord The amber power light on the unit indicates the unit is operating Do not use the unit for at least 5 hours after it is first plugged in P...

Page 6: ...the paraffin and applying consecutive brush strokes to the desired area 4 As an option after final paraffin coating place the coated hand foot or elbow in a Parabath liner or plastic bag and cover wi...

Page 7: ...affin Fig 1 draping over the cover Allow the paraffin to cool and solidify requiring approximately 24 hours Plug the unit in just long enough for the paraffin to loosen in the tank Unplug the unit and...

Page 8: ...t Wax was frozen during shipment Unit was plugged in before it had completely cooled preventing Quick Melt cycle from activating WHAT TO DO Replace the paraffin Replace the paraffin using ParabathPara...

Page 9: ...NO DESCRIPTION 24050 ParabathUnit 110 120V includes 6 lb of Paraffin 24130 Paraffin Refill Unscented 6 lbs 24131 Paraffin Refill Citrus Scent 6 lbs 24212 High Stand 24220 Mitts 24221 Booties 24222 Li...

Page 10: ...uential damages arising out of the proper or improper use prescription or maintenance of this product and in no event shall warrantor s liability for such damages be greater than the purchase price pa...

Page 11: ...apy CLASS I EQUIPMENT TYPE B EQUIPMENT Protective Earth ground ATTENTION READ OWNER S GUIDE 120V 120W 60 Hz MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDA...

Page 12: ...30 633 9359 bonvital com Parabath trademark is property of The Hygenic Corporation and or its subsidiaries and may be registered in the United States and other countries Unauthorized use is strictly p...

Reviews: