Configuration and Calibration Guide HD0679 Rev 1.2.0 27
Unscaled
Calculation
Moisture
Coefficients
Moisture
Calculation
Output Mode
(Current Loop or RS485 request)
Actual Output
Mode F
Mode E
Mode V
A, B, C, D (Mode F)
A, B, C, D (Mode E)
A, B, C, D (Mode V)
Mode F
Mode E
Mode V
Mode F
Mode E
Mode V
Current Loop or
RS485
None
Specified
Selected by
Unscaled 1 Type
(F, E, V)
Figure 12: None specified Output Selection
As older host applications are unable to write the A, B, C and D coefficients for each of the
modes, a final extension is made that supports a set of Legacy Mode coefficients which are
supported by existing host systems. This is shown in the final version of the diagram:
Unscaled
Calculation
Moisture
Coefficients
Moisture
Calculation
Output Mode
(Current Loop or RS485 request)
Actual Output
Mode F
Mode E
Mode V
A, B, C, D (Mode F)
A, B, C, D (Mode E)
A, B, C, D (Mode V)
Mode F
Mode E
Mode V
Mode F
Mode E
Mode V
Current Loop or
RS485
None
Specified
Selected by Unscaled
1 Type (F, E, V)
Selected by
Unscaled 1
Type (F, E, V)
A, B, C, D (Legacy)
Figure 13: Legacy Output Selection
If a current loop output is set without a Mode specifier or an RS485 protocol request is made
without
a Mode specifier
(for a Moisture value) t
hen the following process is followed:
•
If the Legacy Coefficients are non-zero then these are used to calculate the Moisture
value. (Red arrows in the diagram)
•
If the Legacy Coefficients are all zero, then the Unscaled 1 Type is used to select the
relevant coefficients and Moisture (Green arrows). This enables a sensor to be fully
calibrated on a current host system in all modes and operated on a legacy host
system.
9.4
Unscaled 2
In Legacy sensor products a second Unscaled calculation was implemented to allow the
comparison of two modes at the same time. This allowed Unscaled readings for a second
mode to be output, but not Moisture readings. Unscaled 2 has been implemented in the latest
sensors for backwards compatibility, but as these sensors calculate all of the modes all of the
time it should not be used for new host system implementation.
In the latest sensors, multiple RS485 protocol requests can be made to compare modes, or
the two analogue current loop outputs can be configured for different modes.
Summary of Contents for Hydro-Mix
Page 4: ...4 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 6: ...6 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 8: ...8 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 10: ...10 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 14: ...Chapter 1 Introduction 14 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 28: ...Chapter 2 Configuration 28 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 42: ...Chapter 4 Best Practices 42 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 48: ...Chapter 5 Sensor Diagnostics 48 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 52: ...Appendix A Document Cross Reference 52 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...