RS-485 Command
Hydro-Air P-38
24
pH / ORP Analyzer
8.0
RS-485 Command
8.1.
Connections
The controller connects to the RS-485 Modbus RTU via the RS485A (slot 20) and
RS485B (slot 19). Please make sure the connection is correct to the corresponding
slots at the terminal block. See page 6 (Figure 2 and Figure 3).
8.2.
Settings
Assigning ID to the controller
Please ensure the ID assigned in the RS-485 Modbus software is the same ID set in the
controller.
In order to initiate the RS-485 function in the controller, the user needs to set these
configurations in the RS-485 MODBUS software:
CONFIGURATIONS
DETAILS
Protocol
Modbus RTU
Baud rate
9600(Fixed)
Data bits
8
Stop bit
1
Parity
None
ID
0x01 (default)
8.3.
Start Address
ADDRESS
DESCRIPTION
DATA TYPE
DATA
0x00
pH / ORP reading
Sign byte
pH : Reading * 0.01
ORP : Reading * 1
0x01
Current
Byte
Reading * 0.01
0x02
Temperature
Sign byte
Reading * 0.1