Changing the battery
This work must be undertaken in a dry, dust-free environment.
The enclosure
should never be opened inside the hazardous area.
To change the battery on
a Hydro-Logic
®
Flexi Logger 300Ex:
1. Undo the six enclosure screws securing the two halves of the Flexi
Logger 300Ex enclosure using a 2.5m A/F hex driver. The screws are
captive so will stay in the bottom half of the enclosure when undone.
2. Carefully remove the lid of enclosure ensuring that the seals are not lost:
3. Unplug the HLC module and place it to one side:
4. Unplug the battery wire and lift the battery out. Fit the new battery.
5. Replace the HLC module.
6. Refit the enclosure top making sure that the 7 O-ring seals (6 small, one
large) are in place and not contaminated with dirt or other material.
Tighten the six screws to 0.6 N·m, (5in lbf).
7. Reset the logger clock
Resetting the logger clock with
Resetting logger clock with Timeview
Summary of Contents for Hydro-Logic Flexi Logger 300Ex
Page 1: ...www hydro int com LoggerSupport Hydro Logic Flexi Logger 300Ex User Manual ...
Page 66: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 66 of 74 ...
Page 67: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 67 of 74 ...
Page 68: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 68 of 74 ...
Page 69: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 69 of 74 ...
Page 71: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 71 of 74 ...
Page 72: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 72 of 74 ...