User Guide
HYDRAFACIAL
Syndeo
™
1900104-B
en
18.
Do not attempt to lift the unit alone or move over uneven or damaged flooring.
19.
Wearing gloves is recommended when performing all treatments, system cleaning and preventive maintenance activities.
20.
Device cannot be serviced or maintained while in use.
21.
Unauthorized user modifications will void the warranty. HydraFacial, LLC will not be responsible for any injuries
sustained due to unauthorized System modifications or use of parts not specified by HydraFacial, LLC. Systems returned to
HydraFacial, LLC with unsafe modifications will be returned to their original operating condition at the customer’s expense.
22.
DO NOT suction liquids through the dry port located on the side of the unit. This will damage the System and void
the warranty.
23.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
24.
If this System does cause interference, which can be determined by turning the System off and on, the interference can
be possibly corrected by one or more of the following measures:
•
Relocate the System with respect to the receiver.
•
Move the System away from the receiver.
•
Plug the System into a different electrical outlet so that the System and receiver are on different branch circuits.
25.
Use of accessories, transducers, and cables other than those specified or provided by HydraFacial, LLC could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Note:
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 Class A).
If it is used in a residential environment (for which CISPR 11 Class B is normally required), this equipment might not offer adequate
protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
Emissions Standard
Test Method
EN 606061-1-2: 2015
IEC 60601-1-2:2014
CISPR 11:2015 +A1:2016
EN 61000-3-2:2014
EN 61000-3-3:2013
Immunity Standard
Test Method
EN 60601-1-2:2015
IEC 60601-1-2:2014
IEC 61000-4-2:2008
IEC 61000-4-3:2010
IEC 61000-4-4:2012
IEC 61000-4-5:2014
IEC 61000-4-6:2013
IEC 61000-4-8:2009
IEC 61000-4-11:2004
ISO 7637-2
26.
Portable RF communications equipment should be used no closer than 15 cm/6 in to any part of the HydraFacial
System, including cables specified by HydraFacial, LLC. Otherwise, degradation of the performance of this equipment
could result.