9 | Troubleshooting
The following table provides information on troubleshooting:
Malfunction
The reason
Countermeasure
You have no
feeling of
stimulus.
Are the batteries exhausted?
Replace the batteries.
Are the batteries
properly loaded?
Correctly load the
batteries.
Is the wire properly connected? Firmly connect the wire.
Have you removed the
transparent protection film
over the electrode pad?
Remove the transparent
protection film over the
electrode pad.
Stimulus is weak.
Do the electrode pads firmly
stick to the skin?
Firmly stick the electrode pad
to the skin.
Are the electrode pads
overlapped?
Separate the electrode pad
and stick them to the skin gain.
Are the electrode pads dirty?
Please clean the electrode
pad.
Is intensity too weak?
Turn the intensity regulation
dial to regulate it.
Are the electrode pads
properly positioned?
Change the position of the
electrode pad.
The skin
becomes red.
Is the therapeutic time too
long?
Control it within 10 - 15
minutes a time.
Do the electrode pads firmly
stick to the skin?
Please firmly stick the
electrode pad to the skin.
Are the electrode pads too
dry?
Please gently wipe them up
with wet cloth and then use
them again.
Are the electrode pads dirty?
Please clean the electrode
pad.
Are the surface of the
electrode pads scratched?
Please replace them with new
electrode pad.