MCS 1000 Series
Switching logic "active high"
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 44 / 76
BeWa MCS1000 3435427f V130 en-us 2012-06-15.doc
2012-06-15
Switching logic "active high"
Both switching outputs emit a signal when a particle is detected.
You can set the duration of the signal
Δ
t between 5 and 200 ms.
This value is factory set to 7 ms.
The signal gives no indication of the shape and size of the particles.
Switching on the sensor
Note that after the sensor has been switched on or after voltage has returned, a
start-up phase of ~ 70 seconds is necessary before the first measurement can be
made.
Example: Operation - 1 particle detected
0 V
0
U
M1/2
U
t
70 s
Δ
t
3 s