6 | Operation
HYDAC FILTER SYSTEMS GMBH
56 / 124
BeWa CTU1x3x-CTU1x4x 4093420f en-us web
6.3 Selecting the operating modes
Depending on the equipment, the extraction unit provides different
operating modes as described in the following chapters.
WARNING
G60 Spezial or Feroclean L261 - Flammable gas-air
mixture in the analysis chamber
Danger of bodily injury
u
Open fires or smoking are prohibited in the vicinity.
WARNING
Static loading of components
Discharge / Explosion
u
Only use hoses with suitable conductivity.
u
Ground the test item if it is on a support that has an insulating
effect.
u
Only use original spare parts.
CAUTION
Test liquid G60 Spezial or Feroclean L261
Danger to health by spraying the test liquid / aerosols.
u
Keep the door to the analysis chamber closed during the analy-
sis.
6.3.1 Performing the spray extraction
On the user interface, select the spray extraction as the operating
mode.
To spray a test item in a defined manner, use the spray gun (A)
with a specific nozzle (B) in the analysis chamber. Various meter-
ing orifices are available for selection as accessories. For details,
see chapter
.
1
2
3
4
5
6
7
8
9
A