Starting up the module
Starting up the module
CAUTION
Leaking analysis fluid
Danger of eye injury/skin irritation
►
Wear eye protection.
►
Wear protective gloves.
►
Always ensure that all ports that are not in
use are sealed using screw plugs.
Clean-room conditions in the analysis chamber - Flushing of the inner
chamber
Clean the analysis chamber before beginning a contamination test series.
Define the filling volume and flow rate for flushing of the inner chamber.
Always select OUTLET B for this purpose.
Recommended flow rate: 3000 … 4000 ml/min.
Starting/stopping extraction on the CTM-SC module
Press the
key to start extraction.
Press the
key to stop extraction.
Depending on the user interface settings, you may also need to press the
foot switch.
For more details on the user interface, please read the "ConTeS 3xxx"
instructions included in the Technical Documentation for the CTM-SC.