Hycontrol VF05 User And Programming Manual Download Page 13

 

 

13 / 27 

 

5.

 

PROGRAMMING 

There are two ways to program the VF05. 

 

Programming with the HyView software 

 

Programming with the VGF-DISPLAY display unit 
 

5.1.

 

P

ROGRAMMING WITH 

H

Y

V

IEW

 

5.1.1.

 

Installing and Running HyView 

If necessary, install the “

HyView 

HART Configuration Software” (hereafter 

HyView). 

Electrical connections: Connect the transmitter to a PC using a HART

 USB modem (sold separately). 

Start the program and search for the transmitter in the program. 

 

 

5.1.2.

 

Programming and Configuring the Device 

Select the device from the list of found devices to configure and program and open the device's "Device Programming" window. 
 
All the necessary parameters and function settings can be performed with HyView. The table below summarizes the parameters and their location and path. 

 

5.1.2.1

 

Parameters 

Table 1. 

Name (number), function 

Adjustable value range 

Description 

Default value 

Zero-level distance (P04), 
Set container height 

0...60 m 

(0…

200 ft) 

It provides the basis for level measurement calculations. The distance between the plane of the lower 

sealing surface of the mechanical connector (the lower plane of  the flange in the case of a flanged 
version) and the reference point at the bottom of the tank. Its value must be set in a distance unit (Unit 

- P00b). 
Note: There is no measurement beyond the set zero distance if the output is configured for distance or 

level measurement. 

HyView setting: 

Device Settings

” > “Measurement configuration” > “

Zero-

level dist.”

 

VGF-DISPLAY: 

MAIN MENU / CALCULATION / ZERO-LEVEL DISTANCE

 

As per order

 

Minimum measuring distance 

(P05), 

Dead zone 

Warning! 

Critical parameter!

 

Minimum measuring distance… probe 

length (P03) 

The  dead  zone  is  the  distance  between  the  plane  of  the  lower  sealing  surface  of  the  mechanical 

connection (the lower plane of the flange in the case of the flanged version) and the highest level in the 
tank. Signals generated within the dead zone are not processed. The current loop output does not follow 

the dead zone level signal. Its value must be set in distance units (Length unit 

 P00b). 

HyView setting: 

Device Settings

” > “

Measurement 

configuration” >

 

“Minimum”.

 

VGF-DISPLAY: 

MAIN MENU / MEASUREMENT CONFIG / MIN. MEAS. DIST.

 

“Minimum measuring distance” 

according to the Probe data table 

Maximum measuring range limit 

(P06), 

Remote blocking 

0 (limiter off) or the minimum measuring 

distance. The distance between (P05) + 5 

cm and the probe length (P03) 

The maximum value of the measuring range of the device can be limited with this parameter. Signals 
received beyond the set distance will not be processed. Its value is calculated from the plane of the 

lower sealing surface of the mechanical connector (in the case of a flanged version, from the lower 
plane of the flange). 

Remote blocking can be turned off by setting the parameter to 0. In this case, the device will give a 
valid result along the entire length of the sensor. 

HyView setting: 

Device Settings

” > “Measurement optimalization” > “Far end (P06)”.

 

VGF-DISPLAY: 

MAIN MENU / MEASUREMENT CONFIG / MAX. MEAS. DIST.

 

(off) 

Damping time (P20) 

Time constant 

0…999 sec.

 

The "Damping time" is used to reduce undesired fluctuations (e.g.: rippling) in the display of the 

measured data. If the level changes abruptly, the new value will be set to within 1% accuracy 
(exponential setting). Unit of measure: seconds. 

HyView setting: 

Device Settings

” > “Measurement optimalization” > “Damping time”

 

VGF-DISPLAY: 

MAIN MENU / MEAS. OPTIMIZATION / DAMPING TIME

 

10 sec 

Probe length (P03) 

0.

1…30 m

 

(0.33

100 ft) 

The set value is the nominal probe  100 mm (+ 0.33 ft). 
This value should only be changed if the probe length is changed or the probe is replaced. 

Its value must be set in distance unit (Length unit - P00b). For special applications, the length of the 
probe may be greater than the height of the tank but may not exceed 30 m (100 ft). 

HyView setting: 

Device Settings

” > “Measurement configuration” > “Probe length settings”

 

VGF-DISPLAY: 

MAIN MENU / MEASUREMENT CONFIG / PROBE LENGTH

 

Default probe length as per order  

+ 0.1 m (+ 0.33 ft)

 

Fixed output current (P08), 

Setting constant output current 

3.

8…22 mA

 

When the current output is set to "Fixed" mode (P12b = 1), the constant current set here is set. The PV 

current transmission does not work, and at the same time, it overrides any current output fault indication. 

HyView setting:

 

Device Settings

” > “Outputs” > “Fix output current (P8)”

 

VGF-DISPLAY: 

MAIN MENU / OUTPUT SETUP / ANALOG OUTPUT / MANUAL VALUE

 

System of measurement units, 

units of measurement: 

 

Path of parameters:  

Settings

” / Application” 

tab 

Unit system (P00c), 

Unit system settings 

Selectable values: 

 

Metric (EU),  

European unit system 

 

Imperial (US),  
US unit system 

The units that can actually be used (length, volume, weight) change according to the unit system set 
according to this parameter. When setting or changing units, the unit system must be selected first and 

only then can the actual unit to be used be set (according to the previously narrowed list). 

HyView setting:

 

Device Settings

” > “Application” > “Calculation system”

 

VGF-DISPLAY: 

MAIN MENU / BASIC SETUP / UNITS / ENGINEERING SYSTEM

 

Metric (EU) 

Summary of Contents for VF05

Page 1: ...L TRANSMITTERS USER AND PROGRAMMING MANUAL Hycontrol Ltd Larchwood House Orchard Street Redditch Worcestershire UK B98 7DP Tel 44 0 1527 406800 Fax 44 0 1527 406810 Email sales hycontrol com Web www hycontrol com Company registered in England No 1755684 ...

Page 2: ...2 27 LEVEL MEASUREMENT TECHNICAL CONCEPTS ...

Page 3: ... 5 PROGRAMMING 13 5 1 Programming with HyView 13 5 1 1 Installing and Running HyView 13 5 1 2 Programming and Configuring the Device 13 5 1 3 Programming Example 1 Using HyView 18 5 2 Programming with the VGF DISPLAY Display Unit 19 5 2 1 VGF DISPLAY Display Unit 19 5 2 2 The Behavior of the VF05 while Programmed Manually 19 5 3 Properties of VF05 Two Wire Microwave Level Transmitter 21 5 3 1 Leve...

Page 4: ... between the start of the pulse and the arrival of the reflected pulse and the distance traveled is twice the distance to and from the target The device calculates the distance from the measured time into an electric signal proportional to the distance an output current of 4 20 mA and provides a HART output signal and displays it on the display Other derived quantities volume weight etc can also b...

Page 5: ...able gland cable diameter Ø7 13 mm Ø0 3 0 5 or M20x1 5 plastic cable gland cable diameter Ø6 12 mm Ø0 23 0 47 maximum wire cross section 0 5 1 5 mm2 AWG20 AWG15 shielded cable recommended internal thread 2 NPT for the cable s damping tube Electrical protection Class III Weight housing 1 3 kg 2 9 lbs 2 2 kg 4 9 lbs 3 9 kg 8 6 lbs 1 With ideal reflective surfaces and constant temperatures TECHNICAL ...

Page 6: ...robe 6 mm 0 23 12 mm 0 5 16 mm 0 62 Mass 0 16 kg m 0 1 lb feet 0 5 kg m 0 33 lb feet 0 6 kg m 0 4 lb feet Coating of fillet and tension weight PFA PP Tensioning weight dimensions Ø25 x 100 mm Ø1 x 4 Material of tensioning weight 316Ti 1 4571 Maximum medium temperature 150 C 302 F 60 C 140 F MEDIUM TEMPERATURE MEDIUM PRESSURE DIAGRAM Type FLANGE TEMPERATURE 200 C 140 C 80 C 20 C 30 C 1 bar 16 bar 2...

Page 7: ...t of the cast aluminum housing exceeds the limit value so the equipment must be protected against impact and friction 3 4 MAINTENANCE AND REPAIR VF05 does not require maintenance on a regular basis In some very rare instances however the probe may need cleaning to remove deposited material This must be carried out gently without damaging the probe Repairs during or after the warranty period are ca...

Page 8: ...wall the internal anti reflection fittings the rough joints the welds and the shape and design of the top and bottom of the tank are very important for the correct operation of the device Any factor that affects the formation of the electromagnetic field of the probe also significantly impairs the measurement accuracy The coaxial probe is an exception because it has no external electromagnetic fie...

Page 9: ...ensile weight must be straight tensioned and away from other fittings e g mixer They must not come into contact with the tank wall bottom or other objects The installation space requirements of the different measuring probes must be taken into account according to the data of the measuring probes see Technical Data There is no radiation cone along the probe When measuring liquids the use of a refe...

Page 10: ... probe rope depends on the height and shape of the tank the fragmentation of the material the density and the speed of discharge The following table gives the tensile load values for the different materials approximate data in tons Probe type Material Probe length 6 m 20 ft 12 m 40 ft 24 m 80 ft Single cable Ø8 mm Ø0 31 max load 3 0 T Cement 0 6 T 1 2 T 2 4 T Ash 0 3 T 0 6 T 1 2 T The measured pro...

Page 11: ...nst switching harmonics The device is protected against 4 kV ESD Warning The electrostatic discharge protection of the measuring system cannot be solved by the internal ESD protection In all cases it is the user s responsibility to ensure that the tank measured material and probe are grounded Risk of injury The probe may accumulate an electrostatic charge during regular operation so discharge it t...

Page 12: ...4 4 SWITCHING ON AND COMMISSIONING The VF05 two wire device is supplied by HYCONTROL with the technical specifications requested by the customer so it is ready for operation immediately after installation and wiring up Measuring starts less than 20 seconds after switching on Attention The initial current consumption of the device is 3 5 mA right after switching on If there is a change in the insta...

Page 13: ...he measuring range of the device can be limited with this parameter Signals received beyond the set distance will not be processed Its value is calculated from the plane of the lower sealing surface of the mechanical connector in the case of a flanged version from the lower plane of the flange Remote blocking can be turned off by setting the parameter to 0 In this case the device will give a valid...

Page 14: ...de VGF DISPLAY MAIN MENU BASIC SETUP APPLICATION Liquid medium Unit system P00d Temperature unit Selectable values C F The unit of temperature measurement is selected here HyView setting Device Settings Measurement configuration Temperature VGF DISPLAY MAIN MENU BASIC SETUP UNITS ENGINEERING UNITS TEMPERATURE UNITS C Output function P01ab Application mode Selectable values Metric EU Distance Level...

Page 15: ...delay based on the last valid measured data The function is only available for current output with an error signal set to the lower 3 8 mA or the upper 22 mA error current HyView setting Device Settings Measurement optimalization Error delay VGF DISPLAY no adjustment available 0 no delay Special identifying data Path to parameters Device programming window Advanced mode Special tab HART short addr...

Page 16: ... 0 flat 1 2 3 Assigning typical tank bottom designs to a specific tank type for accurate volume calculation The exact form for the setting code is shown in the figure below HyView setting Device settings Tank Silo parameters Bottom shape VGF DISPLAY no adjustment available 0 flat Standing cylindrical tank with dome bottom Standing cylindrical tank with conical bottom Standing rectangular tank Lyin...

Page 17: ... OC table is active When the appropriate changes have been made to the table or it has been filled in correctly click the Send button under the Get button on the right on this page OC table tab to download the table 5 1 2 3 Device status window To turn on the Device status window in HyView right click on the device line in the Device list list in the main window and select Show Device Status Windo...

Page 18: ...beled Send in the lower right row of the window to download the new values to the device 12 Press the X close button to exit the device settings window Creating a volume table HyView OC table OCT A conversion table has to be created using HyView to measure volume The volume values in the table are assigned to the measured levels For asymmetrical and irregularly shaped containers the accuracy of th...

Page 19: ...be damaged If the VF05 cannot be protected against solar radiation not possible or it is used outside the operating temperature range of the VGF DISPLAY do not leave the VGF DISPLAY in the VF05 5 2 2 The Behavior of the VF05 while Programmed Manually By default VF05 displays the main measurement data on the VGF DISPLAY display hereafter referred to as display Enter the programming menu by pressing...

Page 20: ...20 27 Menu hierarchy ...

Page 21: ...gnal with the maximum amplitude that corresponds to the level reflection The following figure is a snapshot of the HyView Echo Chart function typically for a measured material The ADC values of the radar digitizer are located on the vertical axis The ADC value of 4095 corresponds to a radar signal with an amplitude of 3 3V The distance on the horizontal axis 1 Radar reference pulse emitted transmi...

Page 22: ...imum measuring distance 4 The current output range is longer than the measuring range Assignment of 4 mA P10 0 level distance maximum measuring distance Assignment of 20 mA P11 distance longer than level 0 minimum measuring distance 5 The current output range is longer than the measuring range Assignment of 4 mA P10 shorter than 0 level distance maximum measuring distance Assignment of 20 mA P11 d...

Page 23: ...o help differentiate between level echo and interference In case of a weak signal the amplification also helps it highlights the signal from the noise which is easier to measure The automatic gain switching has its own limits Signal gain example automatic by default Gain factor The gain is automatically set by default to make the measured signal as ideal as possible In addition a custom value or g...

Page 24: ...tance The width of the threshold line completion can be adjusted by grasping its extreme points in a similar manner The width of the peaks usually fluctuates less than their height however it is recommended to leave a little on them as well o Deposits may form on the probe the propagation speed of the radar signal in them slows down Thus over time the distance shifts Adding a threshold line positi...

Page 25: ...eck the measurement If no signal is detected change the threshold value manually as described in Threshold line Use the oscilloscope chart and threshold setting functions 0 Reference not found state Occurs when there is a problem with the time base of the measurement card circuit Contact HYCONTROL Level lost and Reference not found state display freezes The probe has been subjected to electrostati...

Page 26: ...ce Programming window Also check that the top layer is at least 100 mm 0 33 ft from the measured material Contact HYCONTROL for the correct setting Event Fault Solution Electrical connection and communication output Current output value 4 mA No power supply Check the power supply Device is not connected properly Check the connection between the device and the power supply The output current settin...

Page 27: ...27 27 June 2022 Hycontrol reserves the right to change technical data without notice ...