10
to the shell of the measured object or the earth. Shield “G” is
connected to the protection parts or the other parts which are
not being measured of the measured object to avoid the error
caused by shell leakage. When measuring the insulation
resistance among components of electronic products, the “L”
terminal and the “E” terminal can be connected to any line.
When measuring the resistance between phases of electrical
generator, you can choose two among the three phases at will,
and the remaining one should be grounded.
V. Notes
1. When the meter is stored you should pay attention to
environment temperature and moisture. The meter should be
place in dry and ventilated environment and should be
shielded from dust, humidity, vibration, acid, soda and
corrosive gas damage.
2. When the measured object is charged, you should
disconnect the power before starting the measurement.
Otherwise, that may endanger safety of your body and the
device! The DC voltage between the “E” terminal and “L”
terminal is high after turning on the high voltage. Make sure
any parts of your body should not touch the terminals during
measuring.