4-1-8. Set tracking motor (Motor Setup)
A. The Motor Setup begins engineer mode.
B. Motor Setup can be set in the type of motor is as follows.
1) X-HEAD
- The optical head portion of the X-axis drive motor
2) Y-BASE
- Y-axis drive motor of the optical base section
3) Z-HEAD
- The optical head portion of the Z-axis drive motor
C. For each of the motors, are the following configurable items.
1) BASE DUTY
- Motor Power of main drive
2) MOVE-LUR
- X, Y, Z axis motor, respectively, left, top, back when driving in
the direction of Power
3) MOVE-RDF
- X, Y, Z axis motor, respectively, right, bottom, front, when driving
in the direction of Power
4) NEAR-LUR
- X, Y, Z axis motor, respectively, left, top, back when you move a
short distance in the direction of Power